3D coronary reconstruction from routine single-plane coronary angiograms: Clinical validation and quantitative analysis of the right coronary artery in 100 patients

被引:70
作者
Messenger, JC
Chen, SYJ
Carroll, JD
Burchenal, JEB
Kioussopoulos, K
Groves, BM
机构
[1] Denver VA Med Ctr, Denver, CO USA
[2] Univ Colorado, Hlth Sci Ctr, Dept Med, Div Cardiol,Adult Cardiac Catheterizat Lab, Denver, CO 80262 USA
来源
INTERNATIONAL JOURNAL OF CARDIAC IMAGING | 2000年 / 16卷 / 06期
关键词
coronary angiography; coronary arteries; image processing (computer-assisted); three-dimensional reconstruction;
D O I
10.1023/A:1010643426720
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Current coronary angiographic techniques display complex three-dimensional (3D) coronary structures in two dimensions (2D). We have developed a 3D reconstruction (3DR) algorithm using standard single-plane angiographic images that allows for 3D display of coronary structures. The purpose of this study was to validate our 3DR algorithm and quantify anatomic characteristics of the right coronary artery (RCA) in vivo. Methods: Accuracy and reproducibility studies were performed using 3DRs of a coronary phantom and in vivo following 3DRs in 40 patients. The anatomic features of the RCA were then quantified in 100 patients. Results: Comparison of length and bifurcation angles (BA) from the phantom to the 3DRs revealed good accuracy and correlation for both (r = 0.95 and 0.93 respectively), with diameter error of <7%. In vivo, the average root mean square (RMS) error in the spatial coordinates of the vessel centerlines was 3.12 +/- 0.77 and 3.16 +/- 0.75 mm in 20 left coronary arteries (LCA) and 20 RCAs respectively. Interobserver average RMS error was 3.47 +/- 1.96 mm and intraobserver average RMS error was 3.02 +/- 1.07 and 3.44 +/- 1.57 mm for two different operators (p = NS). The average RCA length was 10.2 +/- 1.7 cm, average radius of curvature (ROC) was 52 +/- 9 degrees, and the average 3D bifurcation angle of the posterior descending artery (PDA) from the RCA was 55 +/- 22 degrees. Foreshortening (FS) of the segments of the RCA in three 'standard' projections ranged from 0-60, 0-75, and 0-82% respectively. Conclusions: Using our 3DR algorithm patient-specific anatomic characteristics can be accurately displayed and quantified, expanding the information that can be derived from routine coronary angiography.
引用
收藏
页码:413 / 427
页数:15
相关论文
共 31 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Coronary bypass grafts: Assessment of critical 3-D anatomic features [J].
Carroll, JD ;
Chen, SYJ ;
Groves, BM ;
Schaefer, C .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 31 (02) :139A-139A
[3]   3-D reconstruction of coronary arterial tree to optimize angiographic visualization [J].
Chen, SJ ;
Carroll, JD .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2000, 19 (04) :318-336
[4]  
CHEN SY, 1997, CIRCULATION, V96, P290
[5]   Three-dimensional reconstruction of coronary arterial tree based on biplane angiograms [J].
Chen, SYJ ;
Hoffmann, KR ;
Carroll, JD .
MEDICAL IMAGING 1996: IMAGE PROCESSING, 1996, 2710 :103-114
[6]   Improved determination of biplane imaging geometry from two projection images and its application to three-dimensional reconstruction of coronary arterial trees [J].
Chen, SYJ ;
Metz, CE .
MEDICAL PHYSICS, 1997, 24 (05) :633-654
[7]  
Chen SYJ, 1998, DEV CARDIOVASC MED, V204, P61
[8]   Dynamic reconstruction of 3D coronary arterial trees based on a sequence of biplane angiograms [J].
Chen, SYJ ;
Carroll, JD .
IMAGE PROCESSING - MEDICAL IMAGING 1997, PTS 1 AND 2, 1997, 3034 :358-368
[9]  
DELAERE D, 1991, MED BIOL ENG COMPUT, V29, P27
[10]   Dynamics of coronary artery curvature obtained from biplane cineangiograms [J].
Gross, MF ;
Friedman, MH .
JOURNAL OF BIOMECHANICS, 1998, 31 (05) :479-484