Measurement of liver volume using spiral CT and the curved line and cubic spline algorithms: reproducibility and interobserver variation

被引:48
作者
Sandrasegaran, K
Kwo, PW
DiGirolamo, D
Stockberger, SM
Cummings, OW
Kopecky, KK
机构
[1] Indiana Univ Hosp, Dept Radiol, Indianapolis, IN 46202 USA
[2] Indiana Univ Hosp, Dept Med, Indianapolis, IN 46202 USA
[3] Indiana Univ Hosp, Dept Pathol, Indianapolis, IN 46202 USA
来源
ABDOMINAL IMAGING | 1999年 / 24卷 / 01期
关键词
computed tomography (CT); image processing; computer programs; quantitative; liver; CT; volume measurement;
D O I
10.1007/s002619900441
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Background: The purpose of this article was to evaluate the accuracy and reproducibility of spiral computed tomography (CT) and the curved line and cubic spline algorithms in measuring liver volume. Methods: Spiral CT was performed in phantoms, cadaveric liver specimens, and 35 live human subjects (19 healthy volunteers and 16 patients). Images were transferred to a workstation, and volumes were measured by two observers. One observer repeated the measurements at a separate sitting. Results: The correlation between the CT measurement and the gold standard measurement of the cadaveric livers was very strong (r = 0.94). For the live human subjects, the intraobserver and interobserver correlations were extremely high (r = 0.999 and 0.997, respectively). The mean difference in liver volume measurements between the separate observations was 1%. Conclusion: The accuracy and reproducibility of this method of assessing liver volume are very high.
引用
收藏
页码:61 / 65
页数:5
相关论文
共 16 条
[1]   ESTIMATION OF LIVER SIZE BY PERCUSSION IN NORMAL INDIVIDUALS [J].
CASTELL, DO ;
OBRIEN, KD ;
MUENCH, H ;
CHALMERS, TC .
ANNALS OF INTERNAL MEDICINE, 1969, 70 (06) :1183-+
[2]   DETERMINATION OF LIVER VOLUME FROM CT SCANS USING HISTOGRAM CLUSTER-ANALYSIS [J].
FARJO, LA ;
WILLIAMS, DM ;
BLAND, PH ;
FRANCIS, IR ;
MEYER, CR .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1992, 16 (05) :674-683
[3]   Automatic liver segmentation technique for three-dimensional visualisation of CT data [J].
Gao, LM ;
Heath, DG ;
Kuszyk, BS ;
Fishman, EK .
RADIOLOGY, 1996, 201 (02) :359-364
[4]   MEASUREMENT OF LIVER AND SPLEEN VOLUME BY COMPUTED-TOMOGRAPHY - ASSESSMENT OF REPRODUCIBILITY AND CHANGES FOUND FOLLOWING A SELECTIVE DISTAL SPLENORENAL SHUNT [J].
HENDERSON, JM ;
HEYMSFIELD, SB ;
HOROWITZ, J ;
KUTNER, MH .
RADIOLOGY, 1981, 141 (02) :525-527
[5]   ACCURATE MEASUREMENT OF LIVER, KIDNEY, AND SPLEEN VOLUME AND MASS BY COMPUTERIZED AXIAL-TOMOGRAPHY [J].
HEYMSFIELD, SB ;
FULENWIDER, T ;
NORDLINGER, B ;
BARLOW, R ;
SONES, P ;
KUTNER, M .
ANNALS OF INTERNAL MEDICINE, 1979, 90 (02) :185-187
[6]   Fulminant hepatic failure: Observation with serial CT [J].
Itai, Y ;
Sekiyama, K ;
Ahmadi, T ;
Obuchi, M ;
Yoshiba, M .
RADIOLOGY, 1997, 202 (02) :379-382
[7]  
KASAI H, 1992, TRANSPLANTATION, V54, P173
[8]  
KAWASAKI S, 1993, HEPATOLOGY, V18, P1115
[9]   Hepatocyte volume as an indicator of hepatic functional reserve in cirrhotic patients with liver tumours [J].
Matsui, Y ;
Tu, W ;
Kitade, H ;
Nakagawa, A ;
Kamiya, T ;
Kwon, AH ;
Uetsuji, S ;
Kamiyama, Y .
JOURNAL OF GASTROENTEROLOGY AND HEPATOLOGY, 1996, 11 (06) :540-545
[10]   DETERMINATION OF LIVER, KIDNEY, AND SPLEEN VOLUMES BY COMPUTED-TOMOGRAPHY - AN EXPERIMENTAL-STUDY IN DOGS [J].
MOSS, AA ;
FRIEDMAN, MA ;
BRITO, AC .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1981, 5 (01) :12-14