All-polyethylene versus metal-backed and stemmed tibial components in cemented total knee arthroplasty -: A prospective, randomised RSA study

被引:40
作者
Adalberth, G [1 ]
Nilsson, KG
Byström, S
Kolstad, K
Milbrink, J
机构
[1] Univ Uppsala Hosp, Dept Orthopaed Surg, SE-78585 Uppsala, Sweden
[2] Univ Umea Hosp, Dept Orthopaed Surg, SE-90185 Umea, Sweden
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 2001年 / 83B卷 / 06期
关键词
D O I
10.1302/0301-620X.83B6.11430
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We studied the quality of fixation of the tibial component using radiostereometric analysis (RSA) in 40 patients who had undergone a cemented Freeman-Samuelson total knee arthroplasty. They were prospectively randomised to either a stemmed metal-backed (MB) or non-stemmed all-polyethylene (AP) tibial component. The articulating geometry of the implants was identical, as was the operative technique and the postoperative regime. The study showed no complications of fixation using AP tibial components, and the migration was the same as that of their metal-backed counterparts. There was no bony collapse or increased subsidence of any part of the tibial component or increased incidence of radiolucent lines in the knees with AP components. Most AP implants were stable between one and two years after surgery, a finding known to be of positive prognostic significance when predicting future aseptic loosening.
引用
收藏
页码:825 / 831
页数:7
相关论文
共 52 条
[1]
Ahlback S., 1968, Acta Radiol Diagn (Stockh), V277, P7
[2]
THE EFFECT OF A STEM ON THE TIBIAL COMPONENT OF KNEE ARTHROPLASTY - A ROENTGEN STEREOPHOTOGRAMMETRIC STUDY OF UNCEMENTED TIBIAL COMPONENTS IN THE FREEMAN-SAMUELSON KNEE ARTHROPLASTY [J].
ALBREKTSSON, BEJ ;
RYD, L ;
CARLSSON, LV ;
FREEMAN, MAR ;
HERBERTS, P ;
REGNER, L ;
SELVIK, G .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (02) :252-258
[3]
PROXIMALLY CEMENTED VERSUS UNCEMENTED FREEMAN-SAMUELSON KNEE ARTHROPLASTY - A PROSPECTIVE RANDOMIZED STUDY [J].
ALBREKTSSON, BEJ ;
CARLSSON, LV ;
FREEMAN, MAR ;
HERBERTS, P ;
RYD, L .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (02) :233-238
[4]
APEL DM, 1991, CLIN ORTHOP RELAT R, V273, P243
[5]
BARGREN JH, 1983, CLIN ORTHOP RELAT R, P178
[6]
MECHANICAL TESTS ON TIBIAL COMPONENTS OF NON-HINGED KNEE PROSTHESES [J].
BARGREN, JH ;
DAY, WH ;
FREEMAN, MAR ;
SWANSON, SAV .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1978, 60 (02) :256-261
[7]
PERFORMANCE OF THE TIBIAL COMPONENT IN TOTAL KNEE REPLACEMENT - CONVENTIONAL AND REVISION DESIGNS [J].
BARTEL, DL ;
BURSTEIN, AH ;
SANTAVICCA, EA ;
INSALL, JN .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1982, 64 (07) :1026-1033
[8]
THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[9]
THE EFFECT OF CONFORMITY AND PLASTIC THICKNESS ON CONTACT STRESSES IN METAL-BACKED PLASTIC IMPLANTS [J].
BARTEL, DL ;
BURSTEIN, AH ;
TODA, MD ;
EDWARDS, DL .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1985, 107 (03) :193-199
[10]
DORR LD, 1986, CLIN ORTHOP RELAT R, V205, P5