Dosimetric features of linac head and phantom scattered radiation outside the clinical photon beam: experimental measurements and comparison with treatment planning system calculations

被引:30
作者
Cozzi, L [1 ]
Buffa, FM [1 ]
Fogliata, A [1 ]
机构
[1] Ist Oncol Svizzera Italiana, Div Radiat Oncol, Med Phys Unit, Osped S Giovanni, CH-6504 Bellinzona, Switzerland
关键词
high-energy photon beams; contaminating electrons; phantom scatter; head scatter;
D O I
10.1016/S0167-8140(00)00317-0
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: Dosimetric measurements and treatment planning system (TPS) calculations in the region outside the clinical photon beams have been investigated. The aim was to estimate the calculation accuracy of a specific TPS in areas that are becoming increasingly relevant with the advent of new technologies, such as, for example, intensity modulation radiation therapy. Materials and methods: Measurements were performed on two different linacs to obtain, separately, the head scatter (electrons and photons), the transmission below the jaws and the phantom scatter outside the primary beam for different photon energies, distances from the field edge and field sizes. Calculations with a commercial TPS (Helax TMS) were then obtained and compared with these measurements. Results: In general, reasonable agreement between calculations and measurements was obtained (1-2%), especially for photon scattering (head and phantom). Nevertheless, some discrepancies were found in the electron contamination computation, due probably to the approximations and assumptions made in the TPS calculation algorithm. Conclusions: The analyzed TPS presented good results, but for some particular clinical cases and moreover for advanced techniques such as intensity modulated radiation therapy, the calculation behaviour with respect to measurements and patient dose delivery should be carefully evaluated. (C) 2001 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:193 / 200
页数:8
相关论文
共 15 条
[1]   CHARACTERISTICS OF CONTAMINATION ELECTRONS IN HIGH-ENERGY PHOTON BEAMS [J].
BEAUVAIS, H ;
BRIDIER, A ;
DUTREIX, A .
RADIOTHERAPY AND ONCOLOGY, 1993, 29 (03) :308-316
[2]  
Biggs P J, 1979, Med Phys, V6, P291, DOI 10.1118/1.594580
[3]   PERIPHERAL DOSE FROM MEGAVOLT BEAMS [J].
FRAASS, BA ;
VANDEGEIJN, J .
MEDICAL PHYSICS, 1983, 10 (06) :809-818
[4]   DOSE LEVELS OUTSIDE RADIOTHERAPY BEAMS [J].
GREENE, D ;
KARUP, PGG ;
SIMS, C ;
TAYLOR, RJ .
BRITISH JOURNAL OF RADIOLOGY, 1985, 58 (689) :453-456
[5]   DOSE LEVELS OUTSIDE RADIOTHERAPY BEAMS [J].
GREENE, D ;
CHU, GL ;
THOMAS, DW .
BRITISH JOURNAL OF RADIOLOGY, 1983, 56 (668) :543-550
[6]   DOSE IN THE BUILDUP REGION OUTSIDE THE PRIMARY BEAM [J].
HORTON, JL .
MEDICAL PHYSICS, 1984, 11 (03) :331-334
[7]  
IAEA, 1987, IAEA TECHN REP SER, V277
[8]   A STUDY OF THE CHARACTERISTICS OF RADIATION CONTAMINANTS WITHIN A CLINICALLY USEFUL PHOTON-BEAM [J].
KRITHIVAS, G ;
RAO, SN .
MEDICAL PHYSICS, 1985, 12 (06) :764-768
[9]   SOURCES OF ELECTRON CONTAMINATION FOR THE CLINAC-35 25-MV PHOTON-BEAM [J].
PETTI, PL ;
GOODMAN, MS ;
SISTERSON, JM ;
BIGGS, PJ ;
GABRIEL, TA ;
MOHAN, R .
MEDICAL PHYSICS, 1983, 10 (06) :856-861
[10]   INVESTIGATION OF BUILDUP DOSE FROM ELECTRON CONTAMINATION OF CLINICAL PHOTON BEAMS [J].
PETTI, PL ;
GOODMAN, MS ;
GABRIEL, TA ;
MOHAN, R .
MEDICAL PHYSICS, 1983, 10 (01) :18-24