Determination of ultra-trace amounts of arsenic(III) by flow-injection hydride generation atomic absorption spectrometry with on-line preconcentration by coprecipitation with lanthanum hydroxide or hafnium hydroxide

被引:58
作者
Nielsen, S [1 ]
Sloth, JJ [1 ]
Hansen, EH [1 ]
机构
[1] TECH UNIV DENMARK,DEPT CHEM A,DK-2800 LYNGBY,DENMARK
关键词
on-line preconcentration by coprecipitation-dissolution; coprecipitation with hydroxides of lanthanum and hafnium; on-line addition of the coprecipitant; flow-injection hydride generation atomic spectrometry; arsenic(III) assay;
D O I
10.1016/0039-9140(96)01832-2
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A time-based flow-injection (FI) procedure for the determination of ultra-trace amounts of inorganic arsenic(III) is described, which combines hydride generation atomic absorption spectrometry (HG-AAS) with on-line preconcentration of the analyte by inorganic coprecipitation-dissolution in a filterless knotted Microline reactor, The sample and coprecipitating agent are mixed on-line and merged with an ammonium buffer solution, which promotes a controllable and quantitative collection of the generated hydroxide on the inner walls of the knotted reactor incorporated into the FI-HG-AAS system. Subsequently the precipitate is eluted with 1 mol l(-1) hydrochloric acid, allowing ensuing determination of the analyte via hydride generation. The preconcentration of As(III) was tested by coprecipitation with two different inorganic coprecipitating agents namely La(III) and Hf(IV). It was shown that As(III) is more effectively collected by lanthanum hydroxide than by hafnium hydroxide, the sensitivity achieved by the former being approximate to 25% better. With optimal experimental conditions and with a sample consumption of 6.7 mi per assay, an enrichment factor of 32 was obtained at a sample frequency of 33 samples h(-1). The limit of detection (3 sigma) was 0.003 mu g l(-1) and the precision (relative standard deviation) was 1.0% (n = 11) at the 0.1 mu g l(-1) level.
引用
收藏
页码:867 / 880
页数:14
相关论文
共 7 条
[1]  
Fang Z., 1995, FLOW INJECTION ATOMI
[2]  
Fang Z., 1993, FLOW INJECTION SEPAR
[3]   SIMULTANEOUS DETERMINATION OF ARSENIC, BISMUTH, ANTIMONY, SELENIUM AND TELLURIUM IN MOLYBDENUM BY CONTINUOUS HYDRIDE GENERATION AND INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
KUJIRAI, O ;
KOHRI, M ;
YAMADA, K ;
OKOCHI, H .
ANALYTICAL SCIENCES, 1990, 6 (03) :379-383
[4]   Determination of ultra-trace amounts of selenium (IV) by flow injection hydride generation atomic absorption spectrometry with on-line preconcentration by co-precipitation with lanthanum hydroxide .2. On-line addition of co-precipitating agent [J].
Nielsen, S ;
Sloth, JJ ;
Hansen, EH .
ANALYST, 1996, 121 (01) :31-35
[5]   DETERMINATION OF ULTRA-TRACE AMOUNTS OF SELENIUM(IV) BY FLOW-INJECTION HYDRIDE GENERATION ATOMIC-ABSORPTION SPECTROMETRY WITH ONLINE PRECONCENTRATION BY COPRECIPITATION WITH LANTHANUM HYDROXIDE [J].
TAO, GH ;
HANSEN, EH .
ANALYST, 1994, 119 (02) :333-337
[7]   MECHANISMS OF TRANSITION-METAL INTERFERENCES IN HYDRIDE GENERATION ATOMIC-ABSORPTION SPECTROMETRY .1. INFLUENCE OF COBALT, COPPER, IRON AND NICKEL ON SELENIUM DETERMINATION [J].
WELZ, B ;
MELCHER, M .
ANALYST, 1984, 109 (05) :569-572