Real-time accelerated interactive MRI with adaptive TSENSE and UNFOLD

被引:63
作者
Guttman, MA
Kellman, P
Dick, AJ
Lederman, RJ
McVeigh, ER
机构
[1] NHLBI, Cardiac Energet Lab, NIH, Bethesda, MD 20892 USA
[2] NHLBI, Cardiovasc Branch, NIH, Bethesda, MD 20892 USA
关键词
real time; interventional; magnetic resonance imaging; cardiac; adaptive imaging; parallel MRI; SENSE; TSENSE; UNFOLD;
D O I
10.1002/mrm.10504
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Reduced field-of-view (FOV) acceleration using time-adaptive sensitivity encoding (TSENSE) or unaliasing by Fourier encoding the overlaps using the temporal dimension (UNFOLD) can improve the depiction of motion in real-time MRI. However, increased computational resources are required to maintain a high frame rate and low latency in image reconstruction and display. A high-performance software system has been implemented to perform TSENSE and UNFOLD reconstructions for real-time MRI with interactive, on-line display. Images were displayed in the scanner room to investigate image-guided procedures. Examples are shown for normal volunteers and cardiac interventional experiments in animals using a steady-state free precession (SSFP) sequence. In order to maintain adequate image quality for interventional procedures, the imaging rate was limited to seven frames per second after an acceleration factor of 2 with a voxel size of 1.8 x 3.5 x 8 mm. Initial experiences suggest that TSENSE and UNFOLD can each improve the compromise between spatial and temporal resolution in real-time imaging, and can function well in interactive imaging. UNFOLD places no additional constraints on receiver coils, and is therefore more flexible than SENSE methods; however, the temporal image filtering can blur motion and reduce the effective acceleration. Methods are proposed to overcome the challenges presented by the use of TSENSE in interactive imaging. TSENSE may be temporarily disabled after changing the imaging plane to avoid transient artifacts as the sensitivity coefficients adapt. For imaging with a combination of surface and interventional coils, a hybrid reconstruction approach is proposed whereby UNFOLD is used for the interventional coils, and TSENSE with or without UNFOLD is used for the surface coils. Published 2003 Wiley-Liss, Inc.
引用
收藏
页码:315 / 321
页数:7
相关论文
共 26 条
[1]   Multiple field of view MR fluoroscopy [J].
Aksit, P ;
Derbyshire, JA ;
Serfaty, JM ;
Atalar, E .
MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (01) :53-60
[2]  
EGGERS H, 1999, P 7 ANN M ISMRM PHIL, P1649
[3]   Real-time volume rendered MRI for interventional guidance (vol 4, pg 431, 2002) [J].
Guttman, MA ;
Lederman, RJ ;
Sorger, JM ;
McVeigh, ER .
JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2003, 5 (02) :407-407
[4]   Techniques for fast stereoscopic MRI [J].
Guttman, MA ;
McVeigh, ER .
MAGNETIC RESONANCE IN MEDICINE, 2001, 46 (02) :317-323
[5]  
GUTTMAN MA, 2002, P 10 ANN M ISMRM HON, P195
[6]   REAL-TIME ACQUISITION, DISPLAY, AND INTERACTIVE GRAPHIC CONTROL OF NMR CARDIAC PROFILES AND IMAGES [J].
HARDY, CJ ;
DARROW, RD ;
NIETERS, EJ ;
ROEMER, PB ;
WATKINS, RD ;
ADAMS, WJ ;
HATTES, NR ;
MAIER, JK .
MAGNETIC RESONANCE IN MEDICINE, 1993, 29 (05) :667-673
[7]  
Kellman P, 2000, MAGNET RESON MED, V44, P933, DOI 10.1002/1522-2594(200012)44:6<933::AID-MRM15>3.0.CO
[8]  
2-I
[9]   Adaptive sensitivity encoding incorporating temporal filtering (TSENSE) [J].
Kellman, P ;
Epstein, FH ;
McVeigh, ER .
MAGNETIC RESONANCE IN MEDICINE, 2001, 45 (05) :846-852
[10]  
KELLMAN P, 2002, P 10 ANN M ISMRM HON, P857