Use of an enzyme thermistor for semi-continuous blood glucose measurements

被引:7
作者
Carlsson, T [1 ]
Adamson, U [1 ]
Lins, PE [1 ]
Danielsson, B [1 ]
机构
[1] LUND UNIV,CTR CHEM,DEPT PURE & APPL BIOCHEM,S-22100 LUND,SWEDEN
关键词
enzyme thermistor (ET); glucose monitoring; semi-continuous; bedside;
D O I
10.1016/0009-8981(96)06306-1
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
A method to monitor glucose in whole blood is presented. The aim of the project was to develop a prototype for a bedside monitor system for semi-continuous monitoring of the blood glucose concentration, requiring only one calibration. This was made possible by using the special advantage of the thermal sensor technique in combination with the adjustment of flow. The glucose concentration was determined from the difference between the sensor response and an estimated background signal. Using standard addition technique, calibration factors for background and sensitivity were set and remained unchanged during the monitoring. The background signal was 45 +/- 8 mV (mean +/- S.D., n = 8) and the sensitivity was 28 +/- 1 mV/mmol (mean +/- S.D., n = 4). Recovery in whole blood was 90-98% (mean 94%, n = 12). With an injection interval of 3 min the precision with the sensor was <3% over more than 100 blood samples. Response time was about 60 s. The calculated glucose values correlated, r = 0.98, with the values obtained with an YSI glucose analyser (Yellow Springs Instruments, Yellow Springs, OH, USA), over the range 2-20 mmol/l.
引用
收藏
页码:187 / 200
页数:14
相关论文
共 24 条
[1]  
BIN X, 1993, SENS ACTUATORS B, V15, P141
[2]   MICRODIALYSIS OF SUBCUTANEOUS ADIPOSE-TISSUE INVIVO FOR CONTINUOUS GLUCOSE MONITORING IN MAN [J].
BOLINDER, J ;
HAGSTROM, E ;
UNGERSTEDT, U ;
ARNER, P .
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1989, 49 (05) :465-474
[3]  
CLAREMONT DJ, 1989, DIABETOLOGIA, V32, P213
[4]   CALORIMETRIC BIOSENSORS [J].
DANIELSSON, B .
JOURNAL OF BIOTECHNOLOGY, 1990, 15 (03) :187-200
[5]  
DIQUELOU A, 1994, BLOOD, V84, P2206
[6]   DETERMINATION OF GLUCOSE IN DILUTED BLOOD WITH A THERMAL FLOW-INJECTION ANALYSIS BIOSENSOR [J].
HARBORN, U ;
XIE, B ;
DANIELSSON, B .
ANALYTICAL LETTERS, 1994, 27 (14) :2639-2645
[7]  
HARBORN U, IN PRESS EVALUATIONM
[8]  
HASHIGUCHI Y, 1994, DIABETES CARE, V5, P387
[9]   CHARACTERIZATION BY MICRODIALYSIS OF INTERCELLULAR GLUCOSE LEVEL IN SUBCUTANEOUS TISSUE IN HUMANS [J].
JANSSON, PA ;
FOWELIN, J ;
SMITH, U ;
LONNROTH, P .
AMERICAN JOURNAL OF PHYSIOLOGY, 1988, 255 (02) :E218-E220
[10]  
MATTIASSON B, 1976, ANAL LETT, V9, P867, DOI 10.1080/00032717608059154