Acyclic neutral carrier-based polymer membrane electrode for a stimulant, phentermine

被引:10
作者
Katsu, T
Ido, K
Kataoka, K
机构
[1] Okayama Univ, Fac Pharmaceut Sci, Okayama 7008530, Japan
[2] Minist Finance, Cent Customs Lab, Matsudo, Chiba 2710076, Japan
关键词
D O I
10.2116/analsci.17.745
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Groups of dioxadicarboxylic diamides, which were developed as potential ionophores for inorganic cations, were found to act as ionophores for a stimulant, phentermine. Especially, N,N-dioctadecyl-N',N'-dipropyl-3,6-dioxaoctanediamide, which was originally developed as a lead ionophore and is commercially available from Fluka as lead ionophore I, was suitable for making a phentermine-selective electrode. The electrode constructed using this ionophore and bis(2-ethylhexyl) sebacate as a solvent mediator in a poly(vinyl chloride) membrane matrix discriminated between phentermine and analogous compounds more effectively than an electrode based on dibenzo-18-crown-6, a representative ionophore for organic ammonium ions. Moreover, the present electrode showed remarkably little interference by inorganic cations, such as Na+ and K+, as well as lipophilic quaternary ammonium ions including (C2H5)(4)N+ and(C3H7)(4)N+. The electrode exhibited a near-Nernstian response to phentermine in the concentration range of 2 x 10(-6) to 1 x 10(-2) M with a slope of 54.8 mV per concentration decade in 0.1 M MgCl2. The lower limit of detection was 7 x 10(-7) M. This electrode was applied to determine phentermine in a cationic-exchange resin complex of this stimulant, which is the general dosage form in medical use.