Network system for automated seizure detection and contingent delivery of therapy

被引:52
作者
Peters, TE
Bhavaraju, NC
Frei, MG
Osorio, I
机构
[1] Univ Kansas, Med Ctr, Comprehens Epilepsy Ctr, Kansas City, KS 66160 USA
[2] Flint Hills Sci, Lawrence, KS USA
关键词
closed loop; detection; seizure; stimulation;
D O I
10.1097/00004691-200111000-00004
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The authors describe an integrated bedside system for real-time seizure detection and automated delivery of electrical stimulation directly to the brain of subjects undergoing invasive epilepsy surgery evaluation. These stimulations were triggered by specific detections following a prespecified pattern. The system uses a commercially available EEG unit, two personal computers, two Grass S-12 stimulators, and other custorn-built units to enable interfacing between these components. To date, more than 9,500 hours of electrocorticographic data have been acquired, displayed, and analyzed, and more than 900 closed-loop stimulations for seizure blockage have been delivered safely and reliably to eight subjects, with intractable epilepsy. This system generates on-line reports containing information about seizures that provide the epileptologist with timely, valuable data while allowing adaptation of the algorithm detection parameters to improve its performance if necessary. Additionally, it can control the output of any therapeutic device and administrate automatically cognitive tests or radioactive tracers for neuroimaging purposes. This network system, which can be replicated at a relatively low cost by others, is proof of concept for a portable or implantable device that could serve identical functions. Widespread availability of this type of system will advance the fields of clinical and basic epilepsy rapidly and considerably.
引用
收藏
页码:545 / 549
页数:5
相关论文
共 5 条
[1]  
[Anonymous], 1990, NEURAL PROSTHESES FU
[2]  
Diteresi Christopher A., 2000, Epilepsia, V41, P237
[3]   AUTOMATIC RECOGNITION AND QUANTIFICATION OF INTERICTAL EPILEPTIC ACTIVITY IN HUMAN SCALP EEG [J].
GOTMAN, J ;
GLOOR, P .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1976, 41 (05) :513-529
[4]   Real-time automated detection and quantitative analysis of seizures and short-term prediction of clinical onset [J].
Osorio, I ;
Frei, MG ;
Wilkinson, SB .
EPILEPSIA, 1998, 39 (06) :615-627
[5]   An automated injection system (with patient selection) for SPECT imaging in seizure localization [J].
Sepkuty, JP ;
Lesser, TP ;
Civelek, CA ;
Cysyk, B ;
Webber, R ;
Shipley, R .
EPILEPSIA, 1998, 39 (12) :1350-1356