System for the generation of standard gas mixtures of volatile and semi-volatile organic compounds for calibrations of solid-phase microextraction and other sampling devices

被引:67
作者
Koziel, JA [1 ]
Martos, PA [1 ]
Pawliszyn, J [1 ]
机构
[1] Univ Waterloo, Dept Chem, Waterloo, ON N2L 3G1, Canada
关键词
air analysis; gases; volatile organic compounds; semi-volatile organic compounds;
D O I
10.1016/j.chroma.2003.10.079
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Standard gases are used for quality control and quality assurance, development of analysis methods and novel air sampling devices. The use of solid-phase microextraction (SPME) and other novel technologies for research in the area of air sampling and analysis requires systems/devices for reliable standard gas generation and sampling. In this paper we describe a new gas standard generating system for volatile organic compounds (VOCs) and semi-VOCs that was designed, built, and tested to facilitate fundamental and applications research with SPME. The system provided for the generation of a wide range of VOC/semi-VOC concentrations and mixing various standard gases, estimation of detection limits, testing the effects of sampling time, air temperature and relative humidity, testing the effects of air velocity and ozone on sampling/extractions. The system can be also used for calibrations of analytical instrumentation, quality control and quality assurance checks, and cross-validations of SPME with/and other sampling techniques. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:3 / 9
页数:7
相关论文
共 38 条
[1]   Design and validation of portable SPME devices for rapid field air sampling and diffusion based calibration [J].
Augusto, F ;
Koziel, J ;
Pawliszyn, J .
ANALYTICAL CHEMISTRY, 2001, 73 (03) :481-486
[2]  
CAMPBELL E E, 1959, Am Ind Hyg Assoc J, V20, P138, DOI 10.1080/00028895909343685
[3]   Design of an automated analysis system for the determination of organic compounds in continuous air stream using solid-phase microextraction [J].
Eisert, R ;
Pawliszyn, J ;
Barinshteyn, G ;
Chambers, D .
ANALYTICAL COMMUNICATIONS, 1998, 35 (06) :187-189
[4]   NEW METHOD FOR PREPARATION OF PERMEATION TUBES [J].
GODIN, J ;
BOUDENE, C .
ANALYTICA CHIMICA ACTA, 1978, 96 (01) :221-223
[5]   The effect of sample volume on quantitative analysis by solid phase microextraction -: Part 2.: Experimental verification [J].
Górecki, T ;
Khaled, A ;
Pawliszyn, J .
ANALYST, 1998, 123 (12) :2819-2824
[6]   Calibration procedure for solid phase microextraction - gas chromatographic analysis of organic vapours in air [J].
Gorlo, D ;
Wolska, L ;
Zygmunt, B ;
Namiesnik, J .
TALANTA, 1997, 44 (09) :1543-1550
[7]   Solid-phase microextraction for the analysis of human breath [J].
Grote, C ;
Pawliszyn, J .
ANALYTICAL CHEMISTRY, 1997, 69 (04) :587-596
[8]  
HERSCH P, 1964, CHIM ANAL, V46, P31
[9]  
HERSCH P A, 1969, Journal of the Air Pollution Control Association, V19, P164
[10]  
Jia MY, 2000, FIELD ANAL CHEM TECH, V4, P73, DOI 10.1002/1520-6521(2000)4:2/3<73::AID-FACT2>3.0.CO