How accurate is helical CT volumetric assessment in renal tumors?

被引:20
作者
Tann, M [1 ]
Sopov, V
Croitoru, S
Nativ, O
Moskovitz, B
Bar-Meir, E
Groshar, D
机构
[1] Technion Israel Inst Technol, Bnai Zion Med Ctr, Dept Radiol, IL-31048 Haifa, Israel
[2] Technion Israel Inst Technol, Fac Med, IL-31048 Haifa, Israel
[3] Technion Israel Inst Technol, Bnai Zion Med Ctr, Dept Nucl Med, IL-31048 Haifa, Israel
[4] Technion Israel Inst Technol, Bnai Zion Med Ctr, Dept Urol, IL-31048 Haifa, Israel
关键词
CT; measurement; tumor;
D O I
10.1007/s003300000789
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study was to evaluate the accuracy of tumor size measurement on CT studies of renal tumors. Sixteen patients with tumors of the kidneys were imaged by helical CT prior to surgery. Assessment of tumor volume was made by two radiologists on the CT images with the summation of area method, then compared with the resected specimen water displacement volume. Intra- and interobserver agreement for CT measurements were also assessed. There were substantial differences between the CT volume measurement compared with the tumor post-operative volume (mean of differences 30.05 +/- 91.6, 95% CI: 31.45-91.55). The inter- and intraobservation agreements for tumor measurement by CT was found to be satisfactory (ANOVA: p < 0.0001; t-test: p < 0.05). The CT volumetric measurement by area summation is a method with good inter- and intra-examination reproducibility but not an accurate technique for tumor volume assessment.
引用
收藏
页码:1435 / 1438
页数:4
相关论文
共 22 条
[1]  
BALLARD RB, 1995, AM SURGEON, V61, P686
[2]   FACTORS AFFECTING THE MEASUREMENT OF SIZE AND CT-NUMBER IN COMPUTED-TOMOGRAPHY [J].
BAXTER, BS ;
SORENSON, JA .
INVESTIGATIVE RADIOLOGY, 1981, 16 (04) :337-341
[3]   VOLUME DETERMINATIONS USING COMPUTED-TOMOGRAPHY [J].
BREIMAN, RS ;
BECK, JW ;
KOROBKIN, M ;
GLENNY, R ;
AKWARI, OE ;
HEASTON, DK ;
MOORE, AV ;
RAM, PC .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1982, 138 (02) :329-333
[4]  
CERFOLIO RJ, 1993, SURG GYNECOL OBSTET, V176, P307
[5]   Report of an international workshop to standardize response criteria for non-Hodgkin's lymphomas [J].
Cheson, BD ;
Horning, SJ ;
Coiffier, B ;
Shipp, MA ;
Fisher, RI ;
Connors, JM ;
Lister, TA ;
Vose, J ;
Grillo-López, A ;
Hagenbeek, A ;
Cabanillas, F ;
Klippensten, D ;
Hiddemann, W ;
Castellino, R ;
Harris, NL ;
Armitage, JO ;
Carter, W ;
Hoppe, R ;
Canellos, GP .
JOURNAL OF CLINICAL ONCOLOGY, 1999, 17 (04) :1244-1253
[6]   THE ACCURACY OF VOLUMETRIC MEASUREMENT OF HIGH-GRADE GLIOMAS [J].
CHISHOLM, RA ;
STENNING, S ;
HAWKINS, TD .
CLINICAL RADIOLOGY, 1989, 40 (01) :17-21
[7]   ASSESSMENT OF TUMOR RESPONSE BY COMPUTED-TOMOGRAPHY LIVER VOLUMETRY [J].
COSTELLO, P ;
DUSZLAK, EJ ;
LOKICH, J ;
MATELSKI, H ;
CLOUSE, ME .
CT-JOURNAL OF COMPUTED TOMOGRAPHY, 1983, 7 (03) :323-326
[8]   3-DIMENSIONAL QUANTITATION OF PEDIATRIC TUMOR BULK [J].
EGGLI, KD ;
CLOSE, P ;
DILLON, PW ;
UMLAUF, M ;
HOPPER, KD .
PEDIATRIC RADIOLOGY, 1995, 25 (01) :1-6
[9]   TUMOR MEASUREMENT IN THE NUDE-MOUSE [J].
EUHUS, DM ;
HUDD, C ;
LAREGINA, M ;
JOHNSON, FE .
JOURNAL OF SURGICAL ONCOLOGY, 1986, 31 (04) :229-234
[10]   VOLUME AVERAGING LIMITATIONS OF COMPUTED-TOMOGRAPHY [J].
GOODENOUGH, D ;
WEAVER, K ;
DAVIS, D ;
LAFALCE, S .