Impact of Bone Quality and Implant Type on the Primary Stability: An Experimental Study Using Bovine Bone

被引:25
作者
Anil, Sukumaran [1 ]
Aldosari, Abdullah Alfarraj [2 ]
机构
[1] King Saud Univ, Dept Periodont & Community Dent, Coll Dent, Riyadh, Saudi Arabia
[2] King Saud Univ, Dept Prosthet Sci, Dent Implant & Osseointegrat Res Chair, Coll Dent, Riyadh, Saudi Arabia
关键词
implants; removal torque; primary stability; osseointegration; resonance frequency analysis; RESONANCE FREQUENCY-ANALYSIS; FINITE-ELEMENT-ANALYSIS; DENTAL IMPLANTS; BIOMECHANICAL ASPECTS; SURGICAL TECHNIQUE; INSERTION TORQUE; IN-VITRO; OSSEOINTEGRATION; INTERFACE; DENSITY;
D O I
10.1563/AAID-JOI-D-11-00156
中图分类号
R78 [口腔科学];
学科分类号
100302 [口腔临床医学];
摘要
The purpose of this in vitro study was to compare the primary stability and removal torque of bone level and tissue level implants in different bone qualities. Twenty tissue level and bone level implants (3.3 X 10 mm and 4.1 X 10 mm) were used for assessing the stability in type II and type IV bone. Forty bovine rib blocks were used in this study. The primary stability of the implant was measured by the resonance frequency using an Osstel device. The removal torque values (RTV) of the implants was assessed using a digital torque gauge instrument. The implant stability quotient (ISQ) values and the RTV showed a marginally higher stability with bone level implants as compared to tissue level implants. However, these differences were not statistically significant in both type of bone used (P > 0.05). On the other hand, compared to type IV, type II bone showed significant differences in the ISQ (P < 0.01) and RTV (P < 0.001) of bone level and tissue level implants. The study concluded that bone quality is an important factor in establishing primary stability than the implant dimension. Bone level and tissue level implants of same dimensions can be selected based on the esthetic demands since they showed similar mechanical properties.
引用
收藏
页码:144 / 148
页数:5
相关论文
共 54 条
[1]
Implant stability in relation to osseointegration: an experimental study in the Labrador dog [J].
Abrahamsson, Ingemar ;
Linder, Elena ;
Lang, Niklaus P. .
CLINICAL ORAL IMPLANTS RESEARCH, 2009, 20 (03) :313-318
[2]
Akça K, 2007, INT J ORAL MAX IMPL, V22, P101
[3]
Biomechanical aspects of initial intraosseous stability and implant design:: a quantitative micro-morphometric analysis [J].
Akca, Kivanc ;
Chang, Ting-Ling ;
Tekdemir, Ibrahim ;
Fanuscu, Mete I. .
CLINICAL ORAL IMPLANTS RESEARCH, 2006, 17 (04) :465-472
[4]
Study of the osseointegration of dental implants placed with an adapted surgical technique [J].
Al-Marshood, Maysa M. ;
Junker, Rudiger ;
Al-Rasheed, Abdulaziz ;
Aldosari, Abdullah Al Farraj ;
Jansen, John A. ;
Anil, Sukumaran .
CLINICAL ORAL IMPLANTS RESEARCH, 2011, 22 (07) :753-759
[5]
BONE-METAL INTERFACE IN OSSEOINTEGRATION [J].
ALBREKTSSON, T ;
JACOBSSON, M .
JOURNAL OF PROSTHETIC DENTISTRY, 1987, 57 (05) :597-607
[6]
Undersized Implant Site Preparation to Enhance Primary Implant Stability in Poor Bone Density: A Prospective Clinical Study [J].
Alghamdi, Hamdan ;
Anand, Pradeep S. ;
Anil, Sukumaran .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2011, 69 (12) :E506-E512
[7]
Balleri Piero, 2002, Clin Implant Dent Relat Res, V4, P128, DOI 10.1111/j.1708-8208.2002.tb00162.x
[8]
Implant stability measurements for implants placed at the time of extraction: A cohort, prospective clinical trial [J].
Becker, W ;
Sennerby, L ;
Bedrossian, E ;
Becker, BE ;
Lucchini, JP .
JOURNAL OF PERIODONTOLOGY, 2005, 76 (03) :391-397
[9]
Correlation of insertion torques with bone mineral density from dental quantitative CT in the mandible [J].
Beer, A ;
Gahleitner, A ;
Holm, A ;
Tschabitscher, M ;
Homolka, P .
CLINICAL ORAL IMPLANTS RESEARCH, 2003, 14 (05) :616-620
[10]
Primäre Implantatstabilität bei unterschiedlichen KnochenaufbereitungstechnikenEine In-vitro-Untersuchung am Unterkiefer des Minischweins [J].
A. Büchter ;
J. Kleinheinz ;
U. Joos ;
U. Meyer .
Mund-, Kiefer- und Gesichtschirurgie, 2003, 7 (6) :351-355