Power injection of intravenous contrast material through central venous catheters for CT: In vitro evaluation

被引:31
作者
Herts, BR [1 ]
Cohen, MAH [1 ]
McInroy, B [1 ]
Davros, WJ [1 ]
Zepp, RC [1 ]
Einstein, DM [1 ]
机构
[1] CLEVELAND CLIN FDN,DEPT BIOSTAT & EPIDEMIOL,CLEVELAND,OH 44195
关键词
catheters and catheterization; central venous access; technology; contrast media; comparative studies; experimental studies;
D O I
10.1148/radiology.200.3.8756923
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To determine the feasibility of use of a power injector to deliver contrast material through central venous catheters for computed tomographic (CT) examinations. MATERIALS AND METHODS: Ioversol 240 and iothalamate meglumine 43% were separately injected through three 9.6-F Hickman catheters and three 10.0-F Leonard catheters with a power injector in an in vitro study. Flow rates of 1.0, 1.5, 2.0, and 2.5 mL/sec were tested. Peak pressures were mechanically recorded from two sites. A 95% prediction interval was calculated for each peak pressure, and the upper limits at the prediction interval were evaluated to determine if it was less than the recommended limit of 25 psi (175 kPa). RESULTS: Contrast medium, flow rate, and catheter type each statistically significantly affected the measured peak pressures (P = .0001). For each flow rate tested, the upper limits of the prediction interval for the peak pressure at the connection between the coiled tubing and the catheter were below the manufacturer's specified peak pressure. CONCLUSION In vitro analysis demonstrates that power injection of intravenous contrast medium through central venous catheters does not exceed the pressure limits of these catheters at the flow rates tested. In vivo testing to evaluate the safety and efficacy of power injection through central venous catheters is necessary.
引用
收藏
页码:731 / 735
页数:5
相关论文
共 21 条
[1]   CONTRAST MATERIAL FOR COMBINED ABDOMINAL AND PELVIC CT - CAN COST BE REDUCED BY INCREASING THE CONCENTRATION AND DECREASING THE VOLUME [J].
BAKER, ME ;
BEAM, C ;
LEDER, R ;
GULLIVER, D ;
PAINE, SS ;
DUNNICK, NR .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1993, 160 (03) :637-641
[2]   DYNAMIC, INCREMENTAL COMPUTED-TOMOGRAPHY USING A HIGH HEAT-CAPACITY TUBE, SOLID-STATE DETECTOR, AND AUTOMATED INJECTOR POTENTIAL IN BODY IMAGING [J].
BAKER, ME ;
COHAN, R ;
COOPER, C ;
DUNNICK, NR .
CLINICAL IMAGING, 1989, 13 (03) :189-194
[3]   COMPARISON OF IOHEXOL-300 AND DIATRIZOATE MEGLUMINE-60 FOR BODY CT - IMAGE QUALITY, ADVERSE REACTIONS, AND ABORTED REPEATED EXAMINATIONS [J].
BERNARDINO, ME ;
FISHMAN, EK ;
JEFFREY, RB ;
BROWN, PC .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1992, 158 (03) :665-667
[4]   CONTRAST ENHANCEMENT OF HEPATIC-TUMORS IN CT - COMPARISON BETWEEN BOLUS AND INFUSION TECHNIQUES [J].
BURGENER, FA ;
HAMLIN, DJ .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1983, 140 (02) :291-295
[5]   SAFETY CONSIDERATIONS IN THE POWER INJECTION OF CONTRAST-MEDIA VIA CENTRAL VENOUS CATHETERS DURING COMPUTED TOMOGRAPHIC EXAMINATIONS [J].
CARLSON, JE ;
HEDLUND, LJ ;
TRENKNER, SW ;
RITENOUR, R ;
HALVORSEN, RA .
INVESTIGATIVE RADIOLOGY, 1992, 27 (05) :337-340
[6]   BOLUS GEOMETRY AND DYNAMICS AFTER INTRAVENOUS CONTRAST-MEDIUM INJECTION [J].
CLAUSSEN, CD ;
BANZER, D ;
PFRETZSCHNER, C ;
KALENDER, WA ;
SCHORNER, W .
RADIOLOGY, 1984, 153 (02) :365-368
[7]   MAJOR COMPLICATIONS OF PERCUTANEOUS SUBCLAVIAN VEIN CATHETERS [J].
FELICIANO, DV ;
MATTOX, KL ;
GRAHAM, JM ;
BEALL, AC ;
JORDAN, GL .
AMERICAN JOURNAL OF SURGERY, 1979, 138 (06) :869-874
[8]   CONTRAST ENHANCEMENT TECHNIQUE FOR DYNAMIC HEPATIC COMPUTED TOMOGRAPHIC SCANNING [J].
FOLEY, WD ;
BERLAND, LL ;
LAWSON, TL ;
SMITH, DF ;
THORSEN, MK .
RADIOLOGY, 1983, 147 (03) :797-803
[9]   DYNAMIC HEPATIC CT [J].
FOLEY, WD .
RADIOLOGY, 1989, 170 (03) :617-622
[10]   COMPLICATIONS OF INDWELLING CHEMOTHERAPY CATHETERS [J].
GOLDMAN, ML ;
BILBAO, MK ;
ROSCH, J ;
DOTTER, CT .
CANCER, 1975, 36 (06) :1983-1990