Automated multiplex assay system for simultaneous detection of hepatitis B virus DNA, hepatitis C virus RNA, and human immuno deficiency virus type 1 RNA

被引:62
作者
Meng, Q
Wong, C
Rangachari, A
Tamatsukuri, S
Sasaki, M
Fiss, E
Cheng, L
Ramankutty, T
Clarke, D
Yawata, H
Sakakura, Y
Hirose, T
Impraim, C
机构
[1] Roche Mol Syst Inc, Pleasanton, CA 94588 USA
[2] Roche Diagnost KK, Tokyo, Japan
关键词
D O I
10.1128/JCM.39.8.2937-2945.2001
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
We have developed an automated multiplex system for simultaneously screening hepatitis B virus (HBV), hepatitis C virus (HCV), and human immunodeficiency virus type 1 (HIV-1) in blood donations. The assay, designated AMPLINAT MPX HBV/HCV/HIV-1 Test (AMPLINAT MPX), consists of virus extraction and target sequence-specific probe capture on specimen preparation workstation GT-X (Roche Diagnostics K.K., Tokyo, Japan) and amplification and detection by TaqMan PCR on the ABI PRISM 7700 Analyzer (Perkin-Elmer Applied Biosystems, Foster City, Calif.). An internal control (IC) is incorporated in the assay to monitor the extraction, target amplification, and detection processes. The assay yields qualitative results without discrimination of the three targets. Detection limits (95% confidence interval) are 22 to 60 copies/ml for HBV, 61 to 112 IU/ml for HCV, and 33 to 66 copies/ml for HIV-1, using a specimen input volume of 0.2 ml. The AMPLINAT MPX assay detects a broad range of genotypes or subtypes for all three viruses and has a specificity of 99.6% for all three viruses with seronegative specimens. In an evaluation of seroconversion panels, the AMPLINAT MPX assay detects HBV infection an average of 24 days before the detection of HBsAg by enzyme immunoassay. HCV RNA was detected an average of 31 days before HCV antibody. HIV-I RNA was detected an average of 14 days before HIV-1 antibody and an average of 9 days before p24 antigen. The Japanese Red Cross has been evaluating the AMPLINAT MPX system since October 1999. The clinical performance indicates that the AMPLINAT MPX system is robust, sensitive, and reproducible, with a high percentage of valid assay runs (96.8%), a low false-positive rate (0.34%), and a low IC failure rate (0.24%).
引用
收藏
页码:2937 / 2945
页数:9
相关论文
共 21 条
  • [1] Performance of a multiplex qualitative PCR LCx assay for detection of human immunodeficiency virus type 1 (HIV-1) group M subtypes, group O, and HIV-2
    Abravaya, K
    Esping, C
    Hoenle, R
    Gorzowski, J
    Perry, R
    Kroeger, P
    Robinson, J
    Flanders, R
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2000, 38 (02) : 716 - 723
  • [2] DETECTION OF HIV-1 DISTRIBUTION IN DIFFERENT BLOOD FRACTIONS BY 2 NUCLEIC-ACID AMPLIFICATION ASSAYS
    BRUISTEN, S
    VANGEMEN, B
    KOPPELMAN, M
    RASCH, M
    VANSTRIJP, D
    SCHUKKINK, R
    BEYER, R
    WEIGEL, H
    LENS, P
    HUISMAN, H
    [J]. AIDS RESEARCH AND HUMAN RETROVIRUSES, 1993, 9 (03) : 259 - 265
  • [3] TIME-COURSE OF DETECTION OF VIRAL AND SEROLOGIC MARKERS PRECEDING HUMAN-IMMUNODEFICIENCY-VIRUS TYPE-1 SEROCONVERSION - IMPLICATIONS FOR SCREENING OF BLOOD AND TISSUE DONORS
    BUSCH, MP
    LEE, LLL
    SATTEN, GA
    HENRARD, DR
    FARZADEGAN, H
    NELSON, KE
    READ, S
    DODD, RY
    PETERSEN, LR
    [J]. TRANSFUSION, 1995, 35 (02) : 91 - 97
  • [4] HIV AND BLOOD-TRANSFUSIONS - FOCUS ON SEROCONVERSION
    BUSCH, MP
    [J]. VOX SANGUINIS, 1994, 67 : 13 - 18
  • [5] Simultaneous detection of multiplex-amplified human immunodeficiency virus type 1 RNA, hepatitis C virus RNA, and hepatitis B virus DNA using a flow cytometer microsphere-based hybridization assay
    Defoort, JP
    Martin, M
    Casano, B
    Prato, S
    Camilla, C
    Fert, V
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2000, 38 (03) : 1066 - 1071
  • [6] DiDomenico N, 1996, CLIN CHEM, V42, P1915
  • [7] Real time quantitative PCR
    Heid, CA
    Stevens, J
    Livak, KJ
    Williams, PM
    [J]. GENOME RESEARCH, 1996, 6 (10): : 986 - 994
  • [8] Development of a multiplex PCR assay for the simultaneous detection and discrimination of HIV-1, HIV-2, HTLV-I and HTLV-II
    Heredia, A
    Soriano, V
    Weiss, SH
    Bravo, R
    Vallejo, A
    Denny, TN
    Epstein, JS
    Hewlett, IK
    [J]. CLINICAL AND DIAGNOSTIC VIROLOGY, 1996, 7 (02): : 85 - 92
  • [9] SIMULTANEOUS AMPLIFICATION AND DETECTION OF SPECIFIC DNA-SEQUENCES
    HIGUCHI, R
    DOLLINGER, G
    WALSH, PS
    GRIFFITH, R
    [J]. BIO-TECHNOLOGY, 1992, 10 (04): : 413 - 417
  • [10] KINETIC PCR ANALYSIS - REAL-TIME MONITORING OF DNA AMPLIFICATION REACTIONS
    HIGUCHI, R
    FOCKLER, C
    DOLLINGER, G
    WATSON, R
    [J]. BIO-TECHNOLOGY, 1993, 11 (09): : 1026 - 1030