Commutability and traceability:: their repercussions on analytical bias and inaccuracy

被引:12
作者
Ricós, C
Juvany, R
Simón, M
Hernández, A
Alvarez, V
Jiménez, CV
Minchinela, J
Perich, C
机构
[1] Hosp Gen Valle Hebron, Serv Bioquim, E-08035 Barcelona, Spain
[2] Lab Clin Intercomarcal Vilafranca, E-08720 Penedes, Spain
[3] Hosp Barcelona, Clin Lab, E-08901 Barcelona, Spain
[4] Lab Clin Cornella, E-08940 Barcelona, Spain
[5] Lab Barcelones Nord, E-08911 Badalona, Spain
[6] CAP Bon Pastor Lab, E-08030 Barcelona, Spain
关键词
commutability; traceability; biological variation;
D O I
10.1016/S0009-8981(98)00187-9
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
The commutability of calibrators and accuracy control materials affects the traceable link between patient sample results and standards. We sought to identify the repercussions of commutability on various aspects of laboratory practice (calibration, control of bias and accuracy assessment) and to discover the solutions that can reduce the problems produced by noncommutability with presently available resources. Ten serum constituents, ten comparison procedures and 37 analytical procedures were studied. The information concerning accuracy and bias provided from materials found to be commutable in previous works was challenged with native serum results for each routine and reference method compared, using Passing-Bablok regression and decision limits derived from biological variation. We found that: (1) Use of commutable control materials did not assure reliable information on the bias (systematic component of analytical error) of analytical procedures, and (2) Results from native serum and commutable controls were very highly concordant, indicating that these materials provide a good indication of the inaccuracy (total analytical error) of results. We suggest that the performance of individual laboratories would be better evaluated by occasional use of native sera with values assigned by reference methods in EQAS schemes. Moreover, our findings support the idea that manufacturers should assign values to calibrators using reference methods and native sera to reduce matrix effects and promote traceability. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:135 / 145
页数:11
相关论文
共 19 条
[1]  
*BIPM IFC IIFCC IS, 1993, INT VOC BAS GEN TERM
[2]  
Cattozzo G, 1997, EUR J CLIN CHEM CLIN, V35, P387
[3]   MATRIX EFFECTS IN CLINICAL ANALYSIS - COMMUTABILITY OF CONTROL MATERIALS BETWEEN THE EKTACHEM, BECKMAN AND SMA 12-60 GLUCOSE AND UREA METHODS [J].
CLARK, PMS ;
KRICKA, LJ ;
WHITEHEAD, TP .
CLINICA CHIMICA ACTA, 1981, 113 (03) :293-303
[4]  
CORTES M, 1993, BIOQUIMICA CLIN SEMI
[5]  
FRANZINI C, 1993, JIFCC, V5, P186
[6]   ANALYTICAL GOALS FOR THE ACCEPTANCE OF COMMON REFERENCE INTERVALS FOR LABORATORIES THROUGHOUT A GEOGRAPHICAL AREA [J].
GOWANS, EMS ;
PETERSEN, PH ;
BLAABJERG, O ;
HORDER, M .
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1988, 48 (08) :757-764
[7]  
JUVANY R, 1996, THESIS U AUTONOMA BA
[8]  
LASKY FD, 1992, CLIN CHEM, V38, P1260
[9]  
LASKY FD, 1993, ARCH PATHOL LAB MED, V117, P412
[10]  
Libeer JC, 1996, EUR J CLIN CHEM CLIN, V34, P665