Mechanical performance of clinically available, neonatal, high-frequency, oscillatory-type ventilators

被引:40
作者
Hatcher, D
Watanabe, H
Ashbury, T
Vincent, S
Fisher, J
Froese, A
机构
[1] Queens Univ, Dept Anaesthesiol, Kingston, ON, Canada
[2] Queens Univ, Dept Pediat, Kingston, ON K7L 3N6, Canada
[3] Queens Univ, Dept Physiol, Kingston, ON K7L 3N6, Canada
关键词
mechanical ventilation; high-frequency oscillation; neonatal; ventilator output; ventilator frequency dependence; mean airway pressure; monitoring accuracy; waveform;
D O I
10.1097/00003246-199806000-00032
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To perform a functional evaluation of five different high frequency, oscillatory-type ventilators that are currently being marketed for neonatal high-frequency oscillation, Design: Observational animal study, Setting: Laboratory, Subjects: New Zealand White male rabbits, Interventions: Oscillator waveforms and delivered volumes were measured plethysmographically for the following ventilators: the SensorMedics 3100 A; the Drager Baby Log 8000; the Metran Humming V; the Infant Star; and the Infant Star 950, The independent variables which were adjusted included frequency (5 to 15 Hz), amplitude (25% to 100%), mean airway pressure (5 to 25 cm H2O) and lung injury, Measurements and Main Results: At 15 Hz, the volume delivered at the 100% amplitude setting varied from 2.1 to 8.8 mL. Generally, the delivered volume decreased with increasing frequency, and with increased percentage of amplitude. Volume de livery was relatively unaffected by mean airway pressure but decreased with lung injury, Waveforms ranged from pure sinusoidal to a complex square wave. The handling of inspiration/expiration time ratios was ventilator specific, The SensorMedics inspiration/expiration ratio is user selected over a range from 1:2.3 (30% inspiratory time) to 1:1 (50% inspiratory time) and once selected it is consistent over its entire range of operating frequencies, The Drager ratio is machine determined and varied from 1:2.5 at 5 Hz to 1:1 at 15 Hz, Inspiratory time of the Infant Star is machine set at 18 msecs such that the inspiration/expiration ratio is 1:10.1 at 5 Hz and 1:2.7 at 15 Hz, The Humming V has a fixed inspiration/expiration ratio of 1:1, The relationship of the mean airway pressure displayed on the ventilator to the alveolar occlusion pressure varied considerably among devices, The displayed mean pressure could either overestimate (SensorMedics at 33% inspiratory time or Infant Star), approximate (Humming V), or underestimate (Drager) the mean lung distending pressure measured during a brief occlusion maneuver, Conclusions: The findings demonstrate large variations in machine performance. The ventilators also differed profoundly in complexity of operation and versatility as neonatal ventilators.
引用
收藏
页码:1081 / 1088
页数:8
相关论文
共 30 条
[1]   HETEROGENEITY OF MEAN ALVEOLAR PRESSURE DURING HIGH-FREQUENCY OSCILLATIONS [J].
ALLEN, JL ;
FRANTZ, ID ;
FREDBERG, JJ .
JOURNAL OF APPLIED PHYSIOLOGY, 1987, 62 (01) :223-228
[2]   VOLUME RECRUITMENT MANEUVERS ARE LESS DELETERIOUS THAN PERSISTENT LOW LUNG-VOLUMES IN THE ATELECTASIS-PRONE RABBIT LUNG DURING HIGH-FREQUENCY OSCILLATION [J].
BOND, DM ;
FROESE, AB .
CRITICAL CARE MEDICINE, 1993, 21 (03) :402-412
[3]   EFFECT OF MEAN AIRWAY PRESSURE ON GAS-EXCHANGE DURING HIGH-FREQUENCY OSCILLATORY VENTILATION [J].
BOYNTON, BR ;
VILLANUEVA, D ;
HAMMOND, MD ;
VREELAND, PN ;
BUCKLEY, B ;
FRANTZ, ID .
JOURNAL OF APPLIED PHYSIOLOGY, 1991, 70 (02) :701-707
[4]  
BRYAN AC, 1986, AM REV RESPIR DIS, V133, P928
[5]   LUNG-VOLUME RECRUITMENT DURING HIGH-FREQUENCY OSCILLATION IN ATELECTASIS-PRONE RABBITS [J].
BYFORD, LJ ;
FINKLER, JH ;
FROESE, AB .
JOURNAL OF APPLIED PHYSIOLOGY, 1988, 64 (04) :1607-1614
[6]   MECHANISMS OF GAS-TRANSPORT DURING VENTILATION BY HIGH-FREQUENCY OSCILLATION [J].
CHANG, HK .
JOURNAL OF APPLIED PHYSIOLOGY, 1984, 56 (03) :553-563
[7]   RESPIRATORY SYSTEM IMPEDANCE FROM 4 TO 40 HZ IN PARALYZED INTUBATED INFANTS WITH RESPIRATORY-DISEASE [J].
DORKIN, HL ;
STARK, AR ;
WERTHAMMER, JW ;
STRIEDER, DJ ;
FREDBERG, JJ ;
FRANTZ, ID .
JOURNAL OF CLINICAL INVESTIGATION, 1983, 72 (03) :903-910
[8]   ALVEOLAR PRESSURE NONHOMOGENEITY DURING SMALL-AMPLITUDE HIGH-FREQUENCY OSCILLATION [J].
FREDBERG, JJ ;
KEEFE, DH ;
GLASS, GM ;
CASTILE, RG ;
FRANTZ, ID .
JOURNAL OF APPLIED PHYSIOLOGY, 1984, 57 (03) :788-800
[9]   FACTORS INFLUENCING MECHANICAL PERFORMANCE OF NEONATAL HIGH-FREQUENCY VENTILATORS [J].
FREDBERG, JJ ;
GLASS, GM ;
BOYNTON, BR ;
FRANTZ, ID .
JOURNAL OF APPLIED PHYSIOLOGY, 1987, 62 (06) :2485-2490
[10]   ROLE OF LUNG-VOLUME IN LUNG INJURY - HFO IN THE ATELECTASIS-PRONE LUNG [J].
FROESE, AB .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1989, 33 :126-130