Pressure vs flow triggering during pressure support ventilation

被引:26
作者
Goulet, R
Hess, D
Kacmarek, RM
机构
[1] MASSACHUSETTS GEN HOSP,DEPT RESP CARE,BOSTON,MA 02114
[2] HARVARD UNIV,SCH MED,BOSTON,MA
关键词
flow trigger; mechanical ventilation; pressure support ventilation; pressure trigger;
D O I
10.1378/chest.111.6.1649
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: Adult mechanical ventilators have traditionally been pressure- or time-triggered. More recently, flow triggering has become available and some adult ventilators allow the choice between pressure or flow triggering, Prior studies have supported the superiority of flow triggering during continuous positive airway pressure, but few have compared pressure and flow triggering during pressure support ventilation (PSV). The purpose of this study was to compare pressure and flow triggering during PSV in adult mechanically ventilated patients. Methods: The study population consisted of 10 adult patients ventilated with a mechanical ventilator (Nellcor-Puritan-Bennett 7200ae) in the PSV mode. In random order, we compared pressure triggering of -0.5 H2O, pressure triggering -1 cm H2O, flow triggering of 5/2 L/min, and flow triggering 10/3 L/min. Pressure was measured for 5 min at the proximal endotracheal tube using a data acquisition rate of 100 Hz. From the airway pressure signal, trigger pressure (Delta P) was defined as the difference between positive end-expiratory pressure (PEEP) and the maximum negative deflection prior to onset of the triggered breath. Pressure-time product (PTP) was defined as the area produced by the pressure waveform below PEEP during onset of the triggered breath, Trigger time (Delta T) was defined as the time interval below PEEP during onset of the triggered breath. Results: A pressure trigger of -0.5 cm H2O was significantly more sensitive than the other trigger methods for Delta P, PTP, and Delta T (p<0.001), There was also a significant difference between patients for Delta P, Delta T, and PTP for each trigger method (p<0.001). Conclusions: For this group of patients, flow triggering was not superior to pressure triggering at -0.5 cm H2O during PSV.
引用
收藏
页码:1649 / 1653
页数:5
相关论文
共 22 条
[1]   IMPOSED WORK OF BREATHING AND METHODS OF TRIGGERING A DEMAND-FLOW, CONTINUOUS POSITIVE AIRWAY PRESSURE SYSTEM [J].
BANNER, MJ ;
BLANCH, PB ;
KIRBY, RR .
CRITICAL CARE MEDICINE, 1993, 21 (02) :183-190
[2]   COMPONENTS OF THE WORK OF BREATHING AND IMPLICATIONS FOR MONITORING VENTILATOR-DEPENDENT PATIENTS [J].
BANNER, MJ ;
JAEGER, MJ ;
KIRBY, RR .
CRITICAL CARE MEDICINE, 1994, 22 (03) :515-523
[3]  
BHATT SB, 1996, RESP CARE, V41, P512
[4]  
Branson R D, 1994, Respir Care, V39, P138
[5]   COMPARISON OF PRESSURE AND FLOW TRIGGERING SYSTEMS DURING CONTINUOUS POSITIVE AIRWAY PRESSURE [J].
BRANSON, RD ;
CAMPBELL, RS ;
DAVIS, K ;
JOHNSON, DJ .
CHEST, 1994, 106 (02) :540-544
[6]  
BRANSON RD, 1996, RESP CARE, V41, P526
[7]  
CARMACK J, 1994, RESP CARE, V40, pS2
[8]  
Chatburn R L, 1992, Respir Care, V37, P1009
[9]  
GIULIANI R, 1995, AM J RESP CRIT CARE, V151, P1
[10]  
JAGER K, 1994, RESP CARE, V39, P892