Atomic absorption spectroscopy for mercury, automated by sequential injection and miniaturized in lab-on-valve system

被引:200
作者
Erxleben, H [1 ]
Ruzicka, J [1 ]
机构
[1] Univ Washington, Dept Chem, Seattle, WA 98195 USA
关键词
D O I
10.1021/ac058007s
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Sodium borohydride-based hydride generation was automated by using programmable flow within the lab-on-valve module. Mercury vapor, generated in the reaction mixture, was extracted in a gas/liquid separator. The gas-expansion separator was miniaturized and compared with the performance of a novel gas separator that exploits the combination of Venturi effect and reduced pressure. Cold vapor atomic spectroscopy was used as a model system, with detection of mercury by absorption at 254 nm and limit of detection of 9 mu g of Hg/L, using 300 mu L of sample and 100 mu L of borohydride. This work introduces, for the first time, sequential injection technique for hydride generation, highlights advantages of using programmable flow, and outlines means for miniaturization of assays based on spectroscopy of volatile species.
引用
收藏
页码:5124 / 5128
页数:5
相关论文