Clinical Cytometry and Progress in HLA Antibody Detection

被引:35
作者
Bray, Robert A. [1 ]
Tarsitani, Christine [2 ]
Gebel, Howard M. [1 ]
Lee, Jar-How [2 ]
机构
[1] Emory Univ, Dept Pathol, Atlanta, GA 30322 USA
[2] One Lambda Inc, Res Dept, Canoga Pk, CA USA
来源
RECENT ADVANCES IN CYTOMETRY, PART B: ADVANCES IN APPLICATIONS, FIFTH EDITION | 2011年 / 103卷
关键词
RENAL-ALLOGRAFT RECIPIENTS; HUMAN ALLOANTIBODY PROBES; IMPROVED GRAFT-SURVIVAL; CROSS-MATCH TEST; FLOW-CYTOMETRY; KIDNEY-TRANSPLANTATION; CLASS-I; CELL TRANSPLANTATION; REACTIVE ANTIBODIES; MEDIATED REJECTION;
D O I
10.1016/B978-0-12-385493-3.00012-7
中图分类号
Q2 [细胞生物学];
学科分类号
071009 ; 090102 ;
摘要
For most solid organ and selected stem cell transplants, antibodies against mismatched HLA antigens can lead to early and late graft failure. In recognition of the clinical significance of these antibodies, HLA antibody identification is one of the most critical functions of histocompatibility laboratories. Early methods employed cumbersome and insensitive complement-dependent cytotoxicity assays with a visual read-out. A little over 20 years ago flow cytometry entered the realm of antibody detection with the introduction of the flow cytometric crossmatch. Cytometry's increased sensitivity and objectivity quickly earned it popularity as a preferred crossmatch method especially for sensitized recipients. Although a sensitive method, the flow crossmatch was criticized as being "too sensitive" as false positive reactions were a know drawback. In part, the shortcomings of the flow crossmatch were due to the lack of corresponding sensitive and specific HLA antibody screening assays. However, in the mid 1990s, solid phase assays, capable of utilizing standard flow cytometers, were developed. These assays used microparticles coated with purified HLA molecules. Hence, the era of solid-phase, microparticle technology for HLA antibody detection was born permitting the sensitive and specific detection of HLA antibody. It was now possible to provide better correlation between HLA antibody detection and the flow cytometric crossmatch. This flow-based technology was soon followed by adaptation to the Luminex platform permitting a mutltiplexed approach for the identification and characterization of HLA antibodies. It is hoped that these technologies will ultimately lead to the identification of parameters that best correlate with and/or predict transplant outcomes.
引用
收藏
页码:285 / 310
页数:26
相关论文
共 81 条
[1]   A flow cytometric crossmatch test for simultaneous detection of antibodies against donor lymphocytes and endothelial precursor cells [J].
Alheim, M. ;
Johansson, S. M. ;
Hauzenberger, D. ;
Grufman, P. ;
Holgersson, J. .
TISSUE ANTIGENS, 2010, 75 (03) :269-277
[2]   Utility of the virtual crossmatch in solid organ transplantation [J].
Amico, Patrizia ;
Hoenger, Gideon ;
Steiger, Juerg ;
Schaub, Stefan .
CURRENT OPINION IN ORGAN TRANSPLANTATION, 2009, 14 (06) :656-661
[3]  
BERNABEU C, 1983, J IMMUNOL, V131, P2032
[4]   Transplanting the highly sensitized patient: The emory algorithm [J].
Bray, R. A. ;
Nolen, J. D. L. ;
Larsen, C. ;
Pearson, T. ;
Newell, K. A. ;
Kokko, K. ;
Guasch, A. ;
Tso, P. ;
Mendel, J. B. ;
Gebel, H. M. .
AMERICAN JOURNAL OF TRANSPLANTATION, 2006, 6 (10) :2307-2315
[5]   Evolution of HLA antibody detection - Technology emulating biology [J].
Bray, RA ;
Nickerson, PW ;
Kerman, RH ;
Gebel, HM .
IMMUNOLOGIC RESEARCH, 2004, 29 (1-3) :41-53
[6]   THE FLOW CYTOMETRIC CROSSMATCH - DUAL-COLOR ANALYSIS OF T-CELL AND B-CELL REACTIVITIES [J].
BRAY, RA ;
LEBECK, LK ;
GEBEL, HM .
TRANSPLANTATION, 1989, 48 (05) :834-840
[7]  
BRAY RA, 2004, CURRENT PROTOCOL S27, P1
[8]  
BRAY RA, 2007, CLIN FLOW CYTOMETRY, P275
[9]   DETECTION OF PANEL-REACTIVE ANTI-HLA CLASS-I ANTIBODIES BY ENZYME-LINKED-IMMUNOSORBENT-ASSAY OR LYMPHOCYTOTOXICITY - RESULTS OF A BLINDED, CONTROLLED MULTICENTER STUDY [J].
BUELOW, R ;
MERCIER, I ;
GLANVILLE, L ;
REGAN, J ;
ELLINGSON, L ;
JANDA, G ;
CLAAS, F ;
COLOMBE, B ;
GELDER, F ;
GROSSEWILDE, H ;
OROSZ, C ;
WESTHOFF, U ;
VOEGELER, U ;
MONTEIRO, F ;
POULETTY, P .
HUMAN IMMUNOLOGY, 1995, 44 (01) :1-11
[10]   Human leukocyte antigen antibodies for monitoring transplant patients [J].
Cai, JC ;
Terasaki, PI .
SURGERY TODAY, 2005, 35 (08) :605-612