Optimal imaging parameters for fiber-orientation estimation in diffusion MRI

被引:185
作者
Alexander, DC
Barker, GJ
机构
[1] UCL, Dept Comp Sci, London WC1E 6BT, England
[2] Kings Coll London, Dept Neurol, Ctr Neuroimaging Sci, Inst Psychiat, London SE5 8AF, England
关键词
diffusion MRI; Monte Carlo simulation; optimal imaging parameter; crossing fibers; pulse sequence optimization;
D O I
10.1016/j.neuroimage.2005.04.008
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
This study uses Monte Carlo simulations to investigate the optimal value of the diffusion weighting factor b for estimating white-matter fiber orientations using diffusion MRI with a standard spherical sampling scheme. We devise an algorithm for determining the optimal echo time, pulse width, and pulse separation in the pulsed-gradient spinecho sequence for a specific value of b. The Monte Carlo simulations provide an estimate of the optimal value of b for recovering one and two fiber orientations. We show that the optimum is largely independent of the noise level in the measurements and the number of gradient directions and that the optimum depends only weakly on the diffusion anisotropy, the maximum gradient strength, and the spin-spin relaxation time. The optimum depends strongly on the mean diffusivity. In brain tissue, the optima we estimate are in the ranges [0.7, 1.0] x 10(9) s m(-2) and [2.2, 2.81 X 10(9) s m(-2) for the one- and two-fiber cases, respectively. The best b for estimating the fractional anisotropy is slightly higher than for estimating fiber directions in the one-fiber case and slightly lower in the two-fiber case. To estimate Tr(D) in the one-fiber case, the optimal setting is higher still. Simulations suggest that a ratio of high to low b measurements of 5 to I is a good compromise for measuring fiber directions and size and shape indices. (c) 2005 Elsevier Inc. All rights reserved.
引用
收藏
页码:357 / 367
页数:11
相关论文
共 23 条
[1]  
Alexander D. C., 2005, VISUALIZATION IMAGE
[2]  
[Anonymous], PRINCIPLES MAGNETIC
[3]   Utilizing the diffusion-to-noise ratio to optimize magnetic resonance diffusion tensor acquisition strategies for improving measurements of diffusion anisotropy [J].
Armitage, PA ;
Bastin, ME .
MAGNETIC RESONANCE IN MEDICINE, 2001, 45 (06) :1056-1065
[4]   MR DIFFUSION TENSOR SPECTROSCOPY AND IMAGING [J].
BASSER, PJ ;
MATTIELLO, J ;
LEBIHAN, D .
BIOPHYSICAL JOURNAL, 1994, 66 (01) :259-267
[5]  
Basser PJ, 1996, J MAGN RESON SER B, V111, P209, DOI [10.1006/jmrb.1996.0086, 10.1016/j.jmr.2011.09.022]
[6]   Optimization of diffusion measurements using Cramer-Rao lower bound theory and its application to articular cartilage [J].
Brihuega-Moreno, O ;
Heese, FP ;
Hall, LD .
MAGNETIC RESONANCE IN MEDICINE, 2003, 50 (05) :1069-1076
[7]   PULSED-GRADIENT SPIN-ECHO NMR FOR PLANAR, CYLINDRICAL, AND SPHERICAL PORES UNDER CONDITIONS OF WALL RELAXATION [J].
CALLAGHAN, PT .
JOURNAL OF MAGNETIC RESONANCE SERIES A, 1995, 113 (01) :53-59
[8]   Persistent angular structure: new insights from diffusion magnetic resonance imaging data [J].
Jansons, KM ;
Alexander, DC .
INVERSE PROBLEMS, 2003, 19 (05) :1031-1046
[9]   Squashing peanuts and smashing pumpkins: How noise distorts diffusion-weighted MR data [J].
Jones, DK ;
Basser, PJ .
MAGNETIC RESONANCE IN MEDICINE, 2004, 52 (05) :979-993
[10]   Determining and visualizing uncertainty in estimates of fiber orientation from diffusion tensor MRI [J].
Jones, DK .