Propagation of uncertainty in high-performance liquid chromatography with UV-VIS detection

被引:20
作者
Hibbert, DB [1 ]
Jiang, JX
Mulholland, MI
机构
[1] Univ New S Wales, Sch Chem, Sydney, NSW 2052, Australia
[2] Univ Technol Sydney, Dept Appl Chem, Sydney, NSW 2000, Australia
基金
澳大利亚研究理事会;
关键词
uncertainty estimation; error propagation; HPLC; chromatography; Latin Hypercube simulation; UV-VIS detection;
D O I
10.1016/S0003-2670(01)01222-3
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The propagation of uncertainty in peak area or peak height in HPLC with UV-VIS detection, is derived in terms of injection volume, flow rate, injected concentration, retention volume and temperature. It is shown that the sample concentration at the detector has a significant uncertainty distribution, which adds to the uncertainty of the measurement of absorbency. For measurement of peak area and peak height, the uncertainty of the injection volume has the greatest effect. To this is added temperature for peak height, and flow rate and temperature for peak area. Analytical solutions are presented for all uncertainties and are compared with data from a numerical simulation using Latin Hypercube sampling. (C) 2001 Elsevier Science BN. All rights reserved.
引用
收藏
页码:205 / 214
页数:10
相关论文
共 17 条
[1]   DETERMINATION OF THE OPTIMUM WORKING RANGE IN SPECTROPHOTOMETRIC PROCEDURES [J].
ASUERO, AG ;
GONZALEZ, G ;
DEPABLOS, F ;
ARIZA, JLG .
TALANTA, 1988, 35 (07) :531-537
[2]   A Boolean characterization of three-dimensional simple points [J].
Bertrand, G .
PATTERN RECOGNITION LETTERS, 1996, 17 (02) :115-124
[3]  
Draper N., 1981, Applied Regression Analysis, VSecond
[4]   PROPAGATION OF RANDOM ERRORS IN SPECTROPHOTOMETRIC COLORIMETRY [J].
FAIRCHILD, MD ;
RENIFF, L .
COLOR RESEARCH AND APPLICATION, 1991, 16 (06) :360-367
[5]  
HARRIS DC, 1991, QUANTITATIVE CHEM AN
[6]  
HAYASHI Y, 1994, ADV CHROMATOGR, V34, P347
[7]   DEDUCTIVE PREDICTION OF MEASUREMENT PRECISION FROM SIGNAL AND NOISE IN PHOTOMULTIPLIER AND PHOTODIODE DETECTOR OF LIQUID-CHROMATOGRAPHY [J].
HAYASHI, Y ;
MATSUDA, R .
ANALYTICAL SCIENCES, 1994, 10 (05) :725-730
[8]   Measurement precision and 1/f noise in analytical instruments [J].
Hayashi, Y ;
Matsuda, R ;
Poe, RB .
JOURNAL OF CHROMATOGRAPHY A, 1996, 722 (1-2) :157-167
[9]  
Huffel S. V., 1991, TOTAL LEAST SQUARES
[10]  
JONSSON JA, 1987, CHROMATOGRAPHIC SCI, V38, P6