In vitro measurement of patellofemoral force after three types of knee replacement

被引:60
作者
Miller, RK [1 ]
Goodfellow, JW [1 ]
Murray, DW [1 ]
O'Connor, JJ [1 ]
机构
[1] Univ Oxford, Nuffield Orthopaed Ctr, Oxford Orthopaed Engn Ctr, Oxford OX3 7LD, England
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1998年 / 80B卷 / 05期
关键词
D O I
10.1302/0301-620X.80B5.8460
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Using a new, non-invasive method, we measured the patellofemoral force (PFF) in cadaver knees mounted in a rig to simulate weight-bearing. The PFF was measured from 20 degrees to 120 degrees of flexion before and after implanting three designs of knee prosthesis. Medial unicompartmental arthroplasty with a meniscal-bearing prosthesis and with retention of both cruciate ligaments caused no significant change in the PFF. After arthroplasty with a posterior-cruciate-retaining prosthesis and division of the anterior cruciate ligament, the PFF decreased in extension and increased by 20% in flexion. Implantation of a posterior stabilised prosthesis and division of both cruciate ligaments produced a decrease in the PFF in extension but maintained normal load in flexion. There was a direct relationship between the PFF and the angle made with the patellar tendon and the long axis of the tibia. The abnormalities of the patellar tendon angle which resulted from implantation of the two total prostheses explain the observed changes in the PFF and show how the mechanics of the patellofemoral joint depend upon the kinematics of the tibiofemoral articulation.
引用
收藏
页码:900 / 906
页数:7
相关论文
共 24 条
[1]
FORCE ANALYSIS OF THE PATELLAR MECHANISM [J].
AHMED, AM ;
BURKE, DL ;
HYDER, A .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1987, 5 (01) :69-85
[2]
INVITRO MEASUREMENT OF STATIC PRESSURE DISTRIBUTION IN SYNOVIAL JOINTS .2. RETROPATELLAR SURFACE [J].
AHMED, AM ;
BURKE, DL ;
YU, A .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1983, 105 (03) :226-234
[3]
A PRESSURE DISTRIBUTION TRANSDUCER FOR INVITRO STATIC MEASUREMENTS IN SYNOVIAL JOINTS [J].
AHMED, AM .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1983, 105 (03) :309-314
[4]
FAILURE OF METAL-BACKED PATELLAR ARTHROPLASTY - 47 AGC TOTAL KNEES FOLLOWED FOR AT LEAST 1 YEAR [J].
BAECH, J ;
KOFOED, H .
ACTA ORTHOPAEDICA SCANDINAVICA, 1991, 62 (02) :166-168
[5]
FAILURE OF THE METAL-BACKED PATELLAR COMPONENT AFTER TOTAL KNEE REPLACEMENT [J].
BAYLEY, JC ;
SCOTT, RD ;
EWALD, FC ;
HOLMES, GB .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (05) :668-674
[6]
BIDEN E, 1990, KNEE LIGAMENTS STRUC, P135
[7]
A RADIOGRAPHIC STUDY OF BEARING MOVEMENT IN UNICOMPARTMENTAL OXFORD KNEE REPLACEMENTS [J].
BRADLEY, J ;
GOODFELLOW, JW ;
OCONNOR, JJ .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1987, 69 (04) :598-601
[8]
EXPERIMENTAL-DETERMINATION OF FORCES TRANSMITTED THROUGH THE PATELLO-FEMORAL JOINT [J].
BUFF, HU ;
JONES, LC ;
HUNGERFORD, DS .
JOURNAL OF BIOMECHANICS, 1988, 21 (01) :17-23
[9]
Biarticulating two-dimensional computer model of the human patellofemoral joint [J].
Gill, HS ;
OConnor, JJ .
CLINICAL BIOMECHANICS, 1996, 11 (02) :81-89
[10]
THE OXFORD KNEE FOR UNICOMPARTMENTAL OSTEO-ARTHRITIS - THE 1ST 103 CASES [J].
GOODFELLOW, JW ;
KERSHAW, CJ ;
BENSON, MKD ;
OCONNOR, JJ .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1988, 70 (05) :692-701