Collaborative study to establish the first International Standard for quantitation of anti-HPA-1a

被引:22
作者
Allen, D
Rigsby, P
Bessos, H
Berry, J
Wilson, D
Ouwehand, WH
Urbaniak, S
Metcalfe, P
机构
[1] Natl Inst Biol Stand & Controls, Div Haematol, Potters Bar EN6 3QG, Herts, England
[2] Natl Blood Serv, Oxford, England
[3] Natl Inst Biol Stand & Controls, Div Informat, Potters Bar, Herts, England
[4] Scottish Natl Blood Transfus Serv, Edinburgh, Midlothian, Scotland
[5] Scottish Natl Blood Transfus Serv, Aberdeen, Scotland
[6] Univ Cambridge, Dept Haematol, Cambridge, England
[7] Natl Blood Serv, Cambridge, England
关键词
anti-HPA-1a; NAIT; platelet alloantibody detection; quantitation;
D O I
10.1111/j.1423-0410.2005.00669.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background and Objectives This report describes the production of a freeze-dried preparation of pooled human plasma, containing immunoglobulin G (IgG) antibodies against the human platelet antigen 1a (HPA-1a). The material, coded 03/152, is proposed as an International Standard containing 100 arbitrary units of anti-HPA-1a for use in quantitative assays to determine the anti-HPA-1a activity in clinical samples. Materials and Methods Plasma samples containing potent anti-HPA-1a were pooled and freeze dried in 1-ml ampoules. In addition, three individual plasma samples were selected which had varying levels of anti-HPA-1a activity. The anti-HPA-1a activity of these three samples was determined by using a variety of quantitative assays with the proposed standard as a reference. Results An international collaborative study, which was part of the 2004 ISBT Platelet Immunology Workshop, involved 39 laboratories in 24 countries and showed that the anti-HPA-1a activity in three test samples could be reliably determined by using the proposed standard. Conclusions Laboratories can use this standard to measure the anti-HPA-1a activity in patient's samples. Further studies are required to determine the relationship between anti-HPA-1a activity and clinical outcome in patients with neonatal alloimmune thrombocytopenia (NAIT).
引用
收藏
页码:100 / 104
页数:5
相关论文
共 23 条
[1]   Report on the 12th International Society of Blood Transfusion Platelet Immunology Workshop [J].
Bessos, H ;
Wilson, DWL ;
Metcalfe, P ;
Allen, D ;
Urbaniak, SJ .
VOX SANGUINIS, 2005, 89 (02) :105-113
[2]   The development of a quantitative ELISA for antibodies against human platelet antigen type 1a [J].
Bessos, H ;
Perez, S ;
Armstrong-Fisher, S ;
Urbaniak, S ;
Turner, M .
TRANSFUSION, 2003, 43 (03) :350-356
[3]  
BESSOS H, 2004, VOX SANG S3, V87, P42
[4]   European collaborative study of the antenatal management of feto-maternal alloimmune thrombocytopenia [J].
Birchall, JE ;
Murphy, MF ;
Kaplan, C .
BRITISH JOURNAL OF HAEMATOLOGY, 2003, 122 (02) :275-288
[5]   MATERNAL ANTI-D CONCENTRATIONS AND OUTCOME IN RHESUS HEMOLYTIC-DISEASE OF THE NEWBORN [J].
BOWELL, P ;
WAINSCOAT, JS ;
PETO, TEA ;
GUNSON, HH .
BRITISH MEDICAL JOURNAL, 1982, 285 (6338) :327-329
[6]  
Cai ZJ, 1999, WHO TECH REP SER, V887, P1
[7]   INTERNATIONAL BIOLOGICAL STANDARDS AND REFERENCE PREPARATIONS .2. PROCEDURES USED FOR PRODUCTION OF BIOLOGICAL STANDARDS AND REFERENCE PREPARATIONS [J].
CAMPBELL, PJ .
JOURNAL OF BIOLOGICAL STANDARDIZATION, 1974, 2 (04) :259-&
[8]   MONITORING ANTI-HPA-1A PLATELET ANTIBODY-LEVELS DURING PREGNANCY USING THE MAIPA TEST [J].
DAWKINS, B .
VOX SANGUINIS, 1995, 68 (01) :27-34
[9]  
Finney DJ, 1978, STAT METHODS BIOL AS
[10]   Neonatal alloimmune thrombocytopenia due to anti-HPA 1a antibodies;: the level of maternal antibodies predicts the severity of thrombocytopenia in the newborn [J].
Jægtvik, S ;
Husebekk, A ;
Aune, B ;
Oian, P ;
Dahl, LB ;
Skogen, B .
BRITISH JOURNAL OF OBSTETRICS AND GYNAECOLOGY, 2000, 107 (05) :691-694