Portable cell-based biosensor system using integrated CMOS cell-cartridges

被引:134
作者
DeBusschere, BD [1 ]
Kovacs, GTA [1 ]
机构
[1] Stanford Univ, Dept Elect Engn, Stanford, CA 94305 USA
关键词
cell-based biosensor; portable biosensor; CMOS cell-cartridge;
D O I
10.1016/S0956-5663(01)00168-3
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The use of cell-based biosensors outside of the laboratory has been limited due to many issues including preparation of the sample, maintenance of the biological environment, and integration of the electronics for data collection and analysis. This paper describes a system that addresses several of these issues with the development of an integrated silicon-polydimethylsiloxane cell-cartridge. The cell-cartridge contains a CMOS silicon chip that incorporates a digital interface, temperature control system, microelectrode electrophysiology sensors, and analog signal buffering. Additionally, the cell-cartridge supports two separate cell populations in two 10 mul scaled chambers that have independent fluidic channels for sample injection. A portable, microcontroller-based electronics system capable of monitoring the action potential (AP) activity within the cell-cartridges was also developed. The AP activities of cardiomyocyte syncytia in the two chambers differentially responded to the flow of a control medium versus the flow of a biochemical agent. The cell-cartridges and portable electronics system were used to successfully record AP activity from cardiomyocytes outside of the laboratory under realistic application conditions. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:543 / 556
页数:14
相关论文
共 16 条
[1]  
BORKHOLDER DA, 1998, SOL STAT SENS ACT WO, P178
[2]   Whole cell biosensors [J].
Bousse, L .
SENSORS AND ACTUATORS B-CHEMICAL, 1996, 34 (1-3) :270-275
[3]   HL-1 cells: A cardiac muscle cell line that contracts and retains phenotypic characteristics of the adult cardiomyocyte [J].
Claycomb, WC ;
Lanson, NA ;
Stallworth, BS ;
Egeland, DB ;
Delcarpio, JB ;
Bahinski, A ;
Izzo, NJ .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1998, 95 (06) :2979-2984
[4]  
DEBUSSCHERE BD, 1998, SOL STAT SENS ACT C, P358
[5]   Rapid prototyping of microfluidic systems in poly(dimethylsiloxane) [J].
Duffy, DC ;
McDonald, JC ;
Schueller, OJA ;
Whitesides, GM .
ANALYTICAL CHEMISTRY, 1998, 70 (23) :4974-4984
[6]  
Footner P. K., 1987, Quality and Reliability Engineering International, V3, P177, DOI 10.1002/qre.4680030308
[7]  
Freshney RI., 1994, Culture of animal cells : A manual of basic technique, V3rd, DOI DOI 10.1002/9780470649367
[8]  
GILCHRIST KH, BIOS, V16, P557
[9]   TIME DELAYS IN PROPAGATION OF CARDIAC ACTION-POTENTIAL [J].
ISRAEL, DA ;
EDELL, DJ ;
MARK, RG .
AMERICAN JOURNAL OF PHYSIOLOGY, 1990, 258 (06) :H1906-H1917
[10]  
KOVACS GTA, 1990, THESIS STANFORD U ST