Detection of fluorescently labeled microparticles in blood

被引:16
作者
Brandl, M
Hartmann, J
Posnicek, T
Ausenegg, FR
Leitner, A
Falkenhagen, D
机构
[1] Danube Univ Krems, Ctr Biomed Technol, A-3500 Krems, Austria
[2] Christian Doppler Lab Specif Adsorpt Technol Med, Krems, Austria
[3] Karl Franzens Univ Graz, Inst Expt Phys, Graz, Austria
关键词
blood microparticles; extracorporeal circuit; hemofiltration; blood purification; photodetector; microparticles; optical detector; magnetic trap;
D O I
10.1159/000083939
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: A microsphere-based detoxification system is an adsorption system, whereby microadsorbent particles having diameters of 1-20 mu m circulate in an extracorporeal filtrate circle. A thin-wall hollow-fiber membrane filter separates the microparticle-plasma suspension from the bloodstream. For patient safety, it is necessary to have a means to detect membrane ruptures that could lead to a release of microparticles into the patient's bloodstream. Methods: An optical detection system was developed to monitor the venous bloodstream for the presence of microparticles from the filtrate circuit. For detection purposes, cellulose microspheres, both ferromagnetic and fluorescence labeled, were included with the microsphere adsorbant particles. In the case of a membrane rupture, the labeled particles would also be released into the bloodstream. By illuminating a small volume of blood with an excitation wavelength (590 nm) of the fluorescence marker, the particles can be detected by their emission light at 620 nm. The detector sensitivity is increased by collecting the ferromagnetic and fluorescently labeled microparticles using a magnetic trap. The efficiency of magnetic trap arrangements was tested by adjusting the magnet placements. Results: In vitro experiments were performed by pumping whole blood and labeled microparticles through the fluorescence detector. The efficiency of a magnetic trap arrangement was determined. With an optimal trap setup, 5-10 mu l of labeled microparticles can be clearly detected in streaming whole blood. Conclusion: An easy to handle microparticle detector was developed, ready for use in particle based blood detoxification systems. The microparticle detection system fulfills the medical and technical requirements to bring the MDS into clinical tests. Copyright (C) 2005 S. Karger AG, Basel.
引用
收藏
页码:181 / 189
页数:9
相关论文
共 20 条
[1]  
Ash S R, 2000, Ther Apher, V4, P218, DOI 10.1046/j.1526-0968.2000.00192.x
[2]   Treatment of systemic inflammatory response syndrome by push-pull powdered sorbent pheresis: A phase 1 clinical trial [J].
Ash, SR ;
Steczko, J ;
Levy, H ;
Blake, DE ;
Carr, DJ .
THERAPEUTIC APHERESIS, 2001, 5 (06) :497-505
[3]   Techniques and applications - S-layers put to therapeutic use? [J].
Clarkson, S .
NATURE REVIEWS MICROBIOLOGY, 2004, 2 (05) :353-353
[4]  
D Falkenhagen, 1999, US-Patent, Patent No. [5.855.782 /5, 58557825]
[5]   Fractionated plasma separation and adsorption system: A novel system for blood purification to remove albumin bound substances [J].
Falkenhagen, D ;
Strobl, W ;
Vogt, G ;
Schrefl, A ;
Linsberger, I ;
Gerner, FJ ;
Schoenhofen, M .
ARTIFICIAL ORGANS, 1999, 23 (01) :81-86
[6]  
FALKENHAGEN D, 1999, Patent No. 949261697
[7]  
Falkenhagen D., 1995, ASAIO J, V41, P430
[8]  
FALKENHAGEN D, 1999, Patent No. 97501083
[9]  
Hartmann J., 2000, International Journal of Artificial Organs, V23, P555
[10]   SUBPOL: A novel sucrose-based polymer support for solid-phase peptide synthesis and affinity chromatography applications [J].
Poschalko, A ;
Rohr, T ;
Gruber, H ;
Bianco, A ;
Guichard, G ;
Briand, JP ;
Weber, V ;
Falkenhagen, D .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2003, 125 (44) :13415-13426