Ordered subset expectation maximization and wide beam reconstruction "half-time" gated myocardial perfusion SPECT functional imaging: A comparison to "full-time" filtered backprojection

被引:87
作者
DePuey, E. Gordon [1 ,2 ]
Gadiraju, Ramesh [1 ,2 ]
Clark, John [1 ,2 ]
Thompson, Linda [1 ,2 ]
Anstett, Frank [3 ]
Shwartz, Shuli C.
机构
[1] St Lukes Roosevelt Hosp, Div Nucl Med, Dept Radiol, New York, NY 10025 USA
[2] Columbia Univ, Coll Phys & Surg, New York, NY USA
[3] GE Healthcare, Waukesha, WI USA
关键词
D O I
10.1016/j.nuclcard.2008.02.035
中图分类号
R5 [内科学];
学科分类号
1002 [临床医学]; 100201 [内科学];
摘要
Objectives: Compared to filtered back projection (FBP), OSEM with resolution recovery (OSEM-RR) and wide beam reconstruction (WBR)(UltraSPECT Ltd.), which resolve resolution and suppress noise simultaneously during reconstruction, have been shown to maintain/improve myocardial perfusion SPECT quality, even with low count density half-time acquisitions. We postulated that their characteristics would be advantageous for gated SPECT, where each frame is only 1/8th the count density of the summed perfusion images. Methods: An 9 mCi rest/32 mCi (333/1184 MBq) stress Tc-99m sestamibi protocol was used. 15-min FBP, and additional 7-min OSEM-RR and WBR post-stress 8-frame/cardiac cycle SPECT scans were acquired with 90 degrees-angled dual-headed detectors equipped with high resolution collimators in 156 patients. In 82 patients (48F, 34M) (123-252 lbs) with perfusion defects gated image quality was graded visually: 1 (poor)-5 (excellent) Regional LV wall motion (WM) was scored: 0 (normal)-4 (dyskinesis) in a total of 50 vascular territories with defects. Using Myometrix((R)) software (GE Healthcare), post-stress EDV, ESV, and EF were calculated for each method. Additionally, for purposes of comparison, the FBP gated tomograms were processed with other commercially available packages, Emory Toolbox((R)) and Cedars QGS((R)). Results: Despite half-time acquisitions, compared to FBP, image quality increased marginally with OSEM-RR (P = .09) but very significantly with WBR (P = 1.9 x 10(-21)). The WM score was greater only for WBR (P = 4.8 x 10(-8)). Although quantitative parameters correlated well with those determined by FBP (all EF r's > 0.85; all volume r's > 0.93), EFs were significantly lower (P = .0001 for OSEM-RR, 3.4 x 10(-14) for WBR), primarily due to a decrease in EDV with OSEM-RR (P = 7.3 x 10(-13)) and an increase in ESV with WBR (P = 9.2 x 10(-5)). However, inter-method differences in these parameters were of similar magnitude to differences encountered among the commercially available software methods. Conclusions: Half-time OSEM-RR and particularly WBR improve gated SPECT diagnostic quality compared to full-time FBP due to increased resolution and reduced noise. However, these attributes, which affect endocardial edge detection, result in a systematic offset in EDV, ESV, and EF.
引用
收藏
页码:547 / 563
页数:17
相关论文
共 22 条
[1]
Bayesian image reconstruction for emission tomography based on median root prior [J].
Sakari Alenius ;
Ulla Ruotsalainen .
European Journal of Nuclear Medicine, 1997, 24 (3) :258-265
[2]
BASSO D, 2006, J NUCL MED, V47, pP534
[3]
Clinical results of a novel wide beam reconstruction method for shortening scan time of Tc-99m cardiac SPECT perfusion studies [J].
Borges-Neto, Salvador ;
Pagnanelli, Robert A. ;
Shaw, Linda K. ;
Honeycutt, Emily ;
Shwartz, Shuli C. ;
Adams, George L. ;
Coleman, Ralph Edward .
JOURNAL OF NUCLEAR CARDIOLOGY, 2007, 14 (04) :555-565
[4]
Bruyant PP, 2002, J NUCL MED, V43, P1343
[5]
CANTINHO G, 2007, J NUCL CARDIOL, V14, pS56
[6]
DEPUEY E, 1995, J NUCL MED, V36, P952
[7]
DEPUEY EG, 1995, J NUCL MED, V36, P569
[8]
DEPUEY EG, 2006, ANN M AM SOC NUCL CA
[9]
Reversible regional wall motion abnormalities on exercise technetium-99m-gated cardiac single photon emission computed tomography predict high-grade angiographic stenoses [J].
Emmett, L ;
Iwanochko, RM ;
Freeman, MR ;
Barolet, A ;
Lee, DS ;
Husain, M .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (06) :991-998
[10]
Farkash G, 2006, J NUCL CARDIOL, V13, pS5