Third-body wear testing of a highly cross-linked acetabular liner - The effect of large femoral head size in the presence of particulate poly (methyl-methacrylate) debris

被引:26
作者
Bragdon, CR
Jasty, M
Muratoglu, OK
Harris, WH
机构
[1] Massachusetts Gen Hosp, Dept Orthopaed Surg, Orthopaed Biomech & Biomat Lab, Boston, MA 02114 USA
[2] Massachusetts Gen Hosp, Dept Orthopaed Surg, Adult Reconstruct Unit, Boston, MA 02114 USA
[3] Harvard Univ, Sch Med, Boston, MA USA
关键词
total hip arthroplasty; polyethylene; cross-linked; wear; third body; hip simulator;
D O I
10.1016/j.arth.2004.09.035
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The hip simulator wear performance of an electron beam cross-linked and subsequently melted ultrahigh molecular weight polyethylene against femoral heads of 28-, 38-, and 46-mm diameter in the presence of poly (methyl-methacrylate) particulate debris was contrasted with that of conventional polyethylene against a 46-mm diameter head. Over 5 million cycles of testing, the average wear rate of the conventional polyethylene liners was 29.3 +/- 3.0 mg per million cycles. All highly cross-linked components exhibited marked reduction in wear, with the highest wear measuring 0.74 +/- 0.85 mg per million cycles. This study, using a clinically relevant third-body material, showed the electron beam cross-linked material to be far more resistant to this third-body wear than conventional ultrahigh molecular weight polyethylene, even when very large diameter femoral heads were used.
引用
收藏
页码:379 / 385
页数:7
相关论文
共 31 条
[1]
AMSTUTZ H, 1999, 12 ANN INT S TECHN A
[2]
THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[3]
BOBYN JD, 1994, CLIN ORTHOP RELAT R, P27
[4]
Bragdon C R, 1996, Proc Inst Mech Eng H, V210, P157, DOI 10.1243/PIME_PROC_1996_210_408_02
[5]
Bragdon C. R., 2004, J ASTM INT, V1, P1
[6]
Third-body wear of highly cross-linked polyethylene in a hip simulator [J].
Bragdon, CR ;
Jasty, M ;
Muratoglu, OK ;
O'Connor, DO ;
Harris, WH .
JOURNAL OF ARTHROPLASTY, 2003, 18 (05) :553-561
[7]
BRAGDON CR, 1999, 25 ANN M SOC BIOM PR
[8]
Femoral head sizes larger than 32 mm against highly cross-linked polyethylene [J].
Burroughs, BR ;
Rubash, HE ;
Harris, WH .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2002, (405) :150-157
[9]
Larger femoral heads: A triumph of hope over reason! In the affirmative [J].
Callaghan, JJ ;
Brown, TD ;
Pedersen, DR ;
Johnston, RC .
JOURNAL OF ARTHROPLASTY, 2003, 18 (03) :82-84
[10]
Callaghan John J, 2003, Iowa Orthop J, V23, P1