Evaluation of methods for opto-electronic body surface sensing applied to patient position control in breast radiation therapy

被引:20
作者
Baroni, G [1 ]
Troia, A
Riboldi, M
Orecchia, R
Ferrigno, G
Pedotti, A
机构
[1] Politecn Milan, Dept Bioengn, I-20133 Milan, Italy
[2] Politecn Milan, Lab Biomed Technol, I-20133 Milan, Italy
[3] Univ Milan, European Inst Oncol, Div Radiotherapy, Milan, Italy
关键词
radiotherapy; breast cancer; surface registration; opto-electronics; real-time motion analysis;
D O I
10.1007/BF02349976
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
The accuracy gap between the high levels of accuracy in radiotherapy planning and the uncertain set-up of each therapy fraction represents a crucial factor in the optimisation of radiation treatment. This occurs because the conventional means of patient alignment and immobilisation do not guarantee accurate implementation of the therapy plan in the actual irradiation treatment. A patient repositioning technique is proposed, based on opto-electronic motion capture and on methods of registration of body surfaces described by a limited dataset. The validation of the method was related to breast cancer radiotherapy and was based on simulated and experimental repositioning procedures involving a phantom and two subjects. With respect to previous work, the surface registration procedure was, in this case, implemented as a constrained non-linear least-square problem (constraints were given by the position of a couple of passive markers placed on the sternum), and three different algorithms were compared in terms of accuracy in misalignment detection and of computational cost. The simulation and experimental activities identified the best performing algorithm, which systematically limited the repositioning errors to below clinically acceptable thresholds (5mm), with residual surface mismatches lower than 2mm.
引用
收藏
页码:679 / 688
页数:10
相关论文
共 22 条
[1]   Partial surface and volume matching in three dimensions [J].
Barequet, G ;
Sharir, M .
IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1997, 19 (09) :929-948
[2]  
Baroni G, 2000, Comput Aided Surg, V5, P296, DOI 10.3109/10929080009148897
[3]  
Baroni G, 2001, Radiol Med, V102, P168
[4]   Implementation and application of real-time motion analysis based on passive markers [J].
Baroni, G ;
Ferrigno, G ;
Pedotti, A .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1998, 36 (06) :693-703
[5]   Real-time three-dimensional motion analysis for patient positioning verification [J].
Baroni, G ;
Ferrigno, G ;
Orecchia, R ;
Pedotti, A .
RADIOTHERAPY AND ONCOLOGY, 2000, 54 (01) :21-27
[6]  
BARONI G, 2000, Patent No. 00616
[7]   A METHOD FOR REGISTRATION OF 3-D SHAPES [J].
BESL, PJ ;
MCKAY, ND .
IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1992, 14 (02) :239-256
[8]   Autoscan: A flexible and portable 3D scanner [J].
Borghese, NA ;
Ferrigno, G ;
Baroni, G ;
Pedotti, A ;
Ferrari, S ;
Savare, R .
IEEE COMPUTER GRAPHICS AND APPLICATIONS, 1998, 18 (03) :38-41
[9]   The use of an optical outlining system for weekly on-set verification of patients with carcinoma of the breast [J].
Casebow, MP ;
Harris, RL ;
Jarvis, RJ .
RADIOTHERAPY AND ONCOLOGY, 1998, 46 (01) :99-103
[10]   ELITE - A DIGITAL DEDICATED HARDWARE SYSTEM FOR MOVEMENT ANALYSIS VIA REAL-TIME TV SIGNAL-PROCESSING [J].
FERRIGNO, G ;
PEDOTTI, A .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1985, 32 (11) :943-950