The influence of implant position on the wear of alumina-on-alumina studied in a hip simulator

被引:30
作者
Affatato, S
Bersaglia, G
Foltran, I
Emiliani, D
Traina, F
Toni, A
机构
[1] Ist Ortoped Rizzoli, Lab Tecnol Med, I-40136 Bologna, Italy
[2] Div 1 Ortopedia & Traumatol, Bologna, Italy
关键词
hip implant; dislocation; inclination; lubricant; ceramic wear;
D O I
10.1016/S0043-1648(03)00455-1
中图分类号
TH [机械、仪表工业];
学科分类号
0802 [机械工程];
摘要
The most important purpose of total hip arthroplasty is to restore function, relieve pain and obtain stability of the hip joint. The major long term problem in hip replacement surgery is wear of bearing surfaces, the second dislocation of the implant. Besides implant and patient related factors, inappropriate tensional forces due to malposition of acetabular cup, possibly leading to prosthesis impingement, is a main cause for these complications. The incidence of early dislocation following primary total hip arthroplasty is on average between 3 and 5%. Cup malposition is major cause of dislocation, in better cases implant impingement leads to restricted range of motion and higher stresses on bearing surfaces. Therefore having a correct position in the acetabular cups is necessary to reduce dislocation, subluxation, and material failure. Accurate placement of acetabular cups minimizes the risk of implant early dislocations, of high wear rate and the risk of late dislocations. Alternative bearing surfaces, other than polyethylene, have been introduced in the clinical practice to minimize the negative effects of wear on hip arthroplasty survival; an alumina coupling with better tribological properties could be particularly useful when a perfect orientation of implant components is not achievable. In order to evaluate the effect of cup positioning on alumina bearing surfaces, 12 ceramic acetabular cups were tested in a hip joint simulator with bovine calf serum as lubricant. Three configurations were fixed: 23degrees, 45degrees, and 63degrees. After five million cycles, no significant differences were observed among the three tested configurations. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:400 / 405
页数:6
相关论文
共 40 条
[1]
The wear pattern in metal-on-metal hip prostheses [J].
Anissian, HL ;
Stark, A ;
Good, V ;
Dahlstrand, H ;
Clarke, IC .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 2001, 58 (06) :673-678
[2]
[Anonymous], 1979, CLIN ORTHOP RELAT RE
[3]
[Anonymous], 2001, HIP INT
[4]
Ceramic acetabular cups for total hip replacement - Part 6: Cup design, inclination angle and antetorsion angle have an influence on the range of motion and impingement [J].
Bader, R ;
Willmann, G .
BIOMEDIZINISCHE TECHNIK, 1999, 44 (7-8) :212-219
[5]
BADER RJ, 2001, ENG MAT, V192, P549
[6]
BADER RJ, 2001, P 5 C EUR FED NAT AS
[7]
BARDOS DI, 1987, CERAMICS CLIN APPL
[8]
Bono J V, 1994, J Arthroplasty, V9, P119, DOI 10.1016/0883-5403(94)90059-0
[9]
Morphologic characteristics of periprosthetic tissues from hip prostheses with ceramic-ceramic couples - A comparative histologic investigation of 18 revision and 30 autopsy cases [J].
Bos, I ;
Willmann, G .
ACTA ORTHOPAEDICA SCANDINAVICA, 2001, 72 (04) :335-342
[10]
Ultra-low wear rates for rigid-on-rigid bearings in total hip replacements [J].
Clarke, IC ;
Good, V ;
Williams, P ;
Schroeder, D ;
Anissian, L ;
Stark, A ;
Oonishi, H ;
Schuldies, J ;
Gustafson, G .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2000, 214 (H4) :331-347