Incidence and clinical course of thrombus formation on atrial septal defect and patient foramen ovale closure devices in 1,000 consecutive patients

被引:429
作者
Krumsdorf, U [1 ]
Ostermayer, S [1 ]
Billinger, K [1 ]
Trepels, T [1 ]
Zadan, E [1 ]
Horvath, K [1 ]
Sievert, H [1 ]
机构
[1] Cardiovasc Ctr Frankfurt, Frankfurt, Germany
关键词
D O I
10.1016/j.jacc.2003.10.030
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVES The purpose of this study was to investigate the incidence, morphology, and clinical course of thrombus formation after catheter closure of intra-atrial shunts. BACKGROUND Post-procedure detailed information about thrombotic material on different devices for transcatheter closure is missing. METHODS A total of 1,000 consecutive patients were investigated after patent foramen ovale (PFO) (n = 593) or atrial septal defect (ASD) (n = 407) closure. Transesophageal echocardiography (TEE) was scheduled after four weeks and six months. Additional TEEs were performed as clinically indicated. RESULTS Thrombus formation in the left atrium (n = 11), right atrium (n = 6), or both (n = 3) was found in 5 of the 407 (1.2%) ASD patients and in 15 of the 593 (2.5%) PFO patients (p = NS). The thrombus was diagnosed in 14 of 20 patients after four weeks and in 6 of 20 patients later on. The incidence was: 7.1% in the CardioSEAL device (NMT Medical, Boston, Massachusetts); 5.7% in the StarFLEX device (NMT Medical); 6.6% in the PFO-Star device (Applied Biometrics Inc., Burnsville, Minnesota); 3.6% in the ASDOS device (Dr. Ing, Osypka Corp., Grenzach-Wyhlen, Germany); 0.8% in the Helex device (W.L. Gore and Associates, Flagstaff, Arizona); and 0% in the Amplatzer device (AGA Medical Corp., Golden Valley, Minnesota). The difference between the Amplatzer device on one hand and the CardioSEAL device, the StarFLEX device, and the PFO-Star device on the other hand was significant (p < 0.05). A pre-thrombotic disorder as a possible cause of the thrombus was found in two PFO patients. Post-procedure atrial fibrillation (n = 4) and persistent atrial septal aneurysm (n = 4) had been found as significant predictors for thrombus formation (p < 0.05). In 17 of the 20 patients, the thrombus resolved under anticoagulation therapy with heparin or warfarin. In three patients, the thrombus was removed surgically. CONCLUSIONS The incidence of thrombus formation on closure devices is low. The thrombus usually resolves under anticoagulation therapy. (C) 2004 by the American College of Cardiology Foundation.
引用
收藏
页码:302 / 309
页数:8
相关论文
共 40 条
[1]   Transcatheter closure as standard treatment for most interatrial defects:: experience in 200 patients treated with the Amplatzer™ Septal Occluder [J].
Berger, F ;
Ewert, P ;
Björnstad, PG ;
Dähnert, I ;
Krings, G ;
Brilla-Austgenat, I ;
Vogel, M ;
Lange, PE .
CARDIOLOGY IN THE YOUNG, 1999, 9 (05) :468-473
[2]   Transcatheter closure of atrial septal defect and patent foramen ovale in adult patients using the Amplatzer occlusion device: No evidence for thrombus deposition with antiplatelet agents [J].
Brandt, RR ;
Neumann, T ;
Neuzner, J ;
Rau, M ;
Faude, I ;
Hamm, CW .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2002, 15 (10) :1094-1098
[3]   Transcatheter closure of patent foramen ovale in patients with cerebral ischemia [J].
Braun, MU ;
Fassbender, D ;
Schoen, SP ;
Haass, M ;
Schraeder, R ;
Scholtz, W ;
Strasser, RH .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (12) :2019-2025
[4]   Transcatheter closure of atrial septal defect and interatrial communications with a new self expanding nitinol double disc device (Amplatzer septal occluder): multicentre UK experience [J].
Chan, KC ;
Godman, MJ ;
Walsh, K ;
Wilson, N ;
Redington, A ;
Gibbs, JL .
HEART, 1999, 82 (03) :300-306
[5]   Early and late complications associated with transcatheter occlusion of secundum atrial septal defect [J].
Chessa, M ;
Carminati, M ;
Butera, G ;
Bini, RM ;
Drago, M ;
Rosti, L ;
Giamberti, A ;
Pomè, G ;
Bossone, E ;
Frigiola, A .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (06) :1061-1065
[6]  
Cooke J C, 2000, Heart Lung Circ, V9, P30, DOI 10.1046/j.1444-2892.2000.009001030.x
[7]  
Dearani JA, 1999, CIRCULATION, V100, P171
[8]   Prognosis after stroke followed by surgical closure of patent foramen ovale: A prospective follow-up study with brain MRI and simultaneous transesophageal and transcranial Doppler ultrasound [J].
Devuyst, G ;
Bogousslavsky, J ;
Ruchat, P ;
Jeanrenaud, X ;
Despland, AP ;
Regli, F ;
Aebischer, N ;
Karpuz, HM ;
Castillo, V ;
Guffi, M ;
Sadeghi, H .
NEUROLOGY, 1996, 47 (05) :1162-1166
[9]   Comparison between transcatheter and surgical closure of secundum atrial septal defect in children and adults - Results of a multicenter nonrandomized trial [J].
Du, ZD ;
Hijazi, ZM ;
Kleinman, CS ;
Silverman, NH ;
Larnitz, K .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (11) :1836-1844
[10]   Minimally invasive or interventional repair of atrial septal defects in children: Experience in 171 cases and comparison with conventional strategies [J].
Formigari, R ;
Di Donato, RM ;
Mazzera, E ;
Carotti, A ;
Rinelli, G ;
Parisi, F ;
Pasquini, L ;
Ballerini, L .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (06) :1707-1712