Total knee arthroplasty implanted with and without kinematic navigation

被引:89
作者
Hart, R [1 ]
Janecek, M [1 ]
Chaker, A [1 ]
Bucek, P [1 ]
机构
[1] Trauma Hosp, Alloplasty Ctr, Brno 66250, Czech Republic
关键词
D O I
10.1007/s00264-003-0501-6
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Between September 2000 and February 2002 we inserted 120 total knee arthroplasties. In 60 patients we used the standard technique, and in 60 patients we used the OrthoPilot navigation system. Postoperatively all patients had standing long radiographs of the lower extremity from the hip joint to the ankle. We considered the ideal value of the anatomic lateral tibiofemoral angle (LTFA) to be 174degrees. In the standard group the mean value of LTFA was 174.9degrees and in the navigation group 174.3degrees. A deviation between 0degrees and 2degrees from the ideal value was seen in 42 cases in the standard group and in 53 cases in the navigation group. In the standard group 18 cases had a deviation of more than 2.1degrees, whereas there were only seven cases in the navigation group with a deviation exceeding 2.1degrees. There were no complications related to the use of the navigation system. The system affords a possibility to place femoral and tibial components precisely with less axis deviation than with the conventional technique.
引用
收藏
页码:366 / 369
页数:4
相关论文
共 21 条
[1]  
BARGREN JH, 1983, CLIN ORTHOP RELAT R, P178
[2]  
Delp SL, 1998, CLIN ORTHOP RELAT R, P49
[3]   ACCURACY OF INTRAMEDULLARY ALIGNMENT IN TOTAL KNEE REPLACEMENT [J].
ELLOY, MA ;
MANNING, MP ;
JOHNSON, R .
JOURNAL OF BIOMEDICAL ENGINEERING, 1992, 14 (05) :363-370
[4]  
Gosse F, 1997, ORTHOPADE, V26, P258
[5]   CORONAL ALIGNMENT AFTER TOTAL KNEE REPLACEMENT [J].
JEFFERY, RS ;
MORRIS, RW ;
DENHAM, RA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1991, 73 (05) :709-714
[6]   Navigated implantation of total knee prostheses - A comparison with conventional techniques [J].
Jenny, JY ;
Boeri, C .
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 2001, 139 (02) :117-119
[7]   RESULTS OF TOTAL KNEE ARTHROPLASTY AFTER FAILED PROXIMAL TIBIAL OSTEOTOMY FOR OSTEOARTHRITIS [J].
KATZ, MM ;
HUNGERFORD, DS ;
KRACKOW, KA ;
LENNOX, DW .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1987, 69A (02) :225-233
[8]  
Lonner JH, 1996, CLIN ORTHOP RELAT R, P102
[9]   INFLUENCE OF POSITIONING OF PROSTHESIS IN TOTAL KNEE REPLACEMENT [J].
LOTKE, PA ;
ECKER, ML .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1977, 59 (01) :77-79
[10]  
MIELKE RK, 2001, Z ORTHOP, V139, P100