Fast Fourier transform analysis in cavity ring-down spectroscopy:: application to an optical detector for atmospheric NO2

被引:52
作者
Mazurenka, M [1 ]
Wada, R [1 ]
Shillings, AJL [1 ]
Butler, TJA [1 ]
Beames, JM [1 ]
Orr-Ewing, AJ [1 ]
机构
[1] Univ Bristol, Sch Chem, Bristol BS8 1TS, Avon, England
来源
APPLIED PHYSICS B-LASERS AND OPTICS | 2005年 / 81卷 / 01期
关键词
D O I
10.1007/s00340-005-1834-1
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A fast Fourier transform (FFT) method for analysis of ring-down decays from a cavity ring-down (CRD) spectrometer has been tested and compared with alternative fitting methods. The ring-down times derived from the FFT method are obtained with a precision close to that of the Levenberg-Marquardt non-linear least-squares method, but the fitting algorithm is similar to 100 times faster, allowing real-time fitting of individual ring-down traces on a personal computer. Advantages of the FFT method are discussed, and the method is demonstrated for the measurement of NO2 partial pressures equivalent to mixing ratios of 150 pptv and above in laboratory air, using a CRD spectrometer based on an external cavity diode laser operating at wavelengths around 410 nm. The absorption by NO2 is distinguished from other cavity losses either by using synthetic (zero) air reference samples, or by diverting the sampled laboratory airflow through an NO2 chemical scrubber consisting of hydroxyapatite on a TiO2 substrate. Typical mixing ratios of NO2 in the laboratory air are similar to 25 ppbv.
引用
收藏
页码:135 / 141
页数:7
相关论文
共 34 条
[1]   Solving chemical problems of environmental importance using cavity ring-down spectroscopy [J].
Atkinson, DB .
ANALYST, 2003, 128 (02) :117-125
[2]   Sensitive absorption measurements in the near-infrared region using off-axis integrated-cavity-output spectroscopy [J].
Baer, DS ;
Paul, JB ;
Gupta, JB ;
O'Keefe, A .
APPLIED PHYSICS B-LASERS AND OPTICS, 2002, 75 (2-3) :261-265
[3]  
BALL CAB, 1992, MATER SCI FORUM, V83, P1147, DOI 10.4028/www.scientific.net/MSF.83-87.1147
[4]   Broad-band cavity ring-down spectroscopy [J].
Ball, SM ;
Jones, RL .
CHEMICAL REVIEWS, 2003, 103 (12) :5239-5262
[5]   Cavity ring-down spectroscopy: Experimental schemes and applications [J].
Berden, G ;
Peeters, R ;
Meijer, G .
INTERNATIONAL REVIEWS IN PHYSICAL CHEMISTRY, 2000, 19 (04) :565-607
[6]   Absorption spectroscopy in high-finesse cavities for atmospheric studies [J].
Brown, SS .
CHEMICAL REVIEWS, 2003, 103 (12) :5219-5238
[7]   Simultaneous in situ detection of atmospheric NO3 and N2O5 via cavity ring-down spectroscopy [J].
Brown, SS ;
Stark, H ;
Ciciora, SJ ;
McLaughlin, RJ ;
Ravishankara, AR .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2002, 73 (09) :3291-3301
[8]   APPLICATION OF THE DIVISOR METHOD TO MULTIPLE PEAK DLTS SPECTRA [J].
DEVRIES, PD ;
KHAN, AA .
JOURNAL OF ELECTRONIC MATERIALS, 1989, 18 (06) :763-766
[9]   AN EFFICIENT TECHNIQUE FOR ANALYZING DEEP LEVEL TRANSIENT SPECTROSCOPY DATA [J].
DEVRIES, PD ;
KHAN, AA .
JOURNAL OF ELECTRONIC MATERIALS, 1989, 18 (04) :543-547
[10]   ERRORS CAUSED BY BASE-LINE OFFSET AND NOISE IN THE ESTIMATION OF EXPONENTIAL LIFETIMES [J].
DOWELL, LJ ;
GILLIES, GT .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (01) :242-243