High pacing impedances: Are you overtorquing your leads?

被引:17
作者
Laske, TG
Vieau, SA
Skadsberg, ND
Iaizzo, PA
机构
[1] Univ Minnesota, Dept Surg, Minneapolis, MN 55455 USA
[2] Univ Minnesota, Dept Biomed Engn, Minneapolis, MN 55455 USA
[3] Univ Minnesota, Dept Physiol, Minneapolis, MN 55455 USA
[4] Medtronic Inc, Minneapolis, MN USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2005年 / 28卷 / 09期
关键词
atrium; electrophysiology; endocardium; impedance; pacing;
D O I
10.1111/j.1540-8159.2005.00192.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
High Pacing Impedances: Are You Overtorquing Your Leads? Background: Variations in measured pacing impedances that occur at the time of lead implantation remain largely unexplained and may be due to the morphology of the tissue-leod interface. Methods: An endocordial pacing lead was implanted under direct endoscopic visualization and parameters were measured for defined stages of implantation into multiple sites within the right atrium of in vitro swine hearts (n = 6, 38 implants), in vivo swine hearts (n = 2, 10 implants), and an in vitro human heart (n = 1, 15 implants). Results: Steady increases in impedance values up to 2 turns fully fixed (2TF) were associated with minimal tissue distortion in all implants. Overtorquing of the in vitro swine implants resulted in severe distortion at the tissue-lead interface demonstrating either tissue wrapping (24 implants) or tissue coring (14 implants). Impedance and threshold values remained elevated (953 +/- 282 Omega, 7.86 +/- 3.0 V; both P < 0.05 vs 2TF) during tissue distortion/wrapping, while tissue-cored implants were associated with significant decreases (552 +/- 187 Omega, 6.2 +/- 2.2 V, both P < 0.05 vs 2TF). P-wave amplitudes demonstrated no significant changes or correlation to tissue distortion. Importantly, both swine in vivo and human in vitro data demonstrated similar trends compared with the swine in vitro data. Conclusions: In this study, one is able to directly observe and correlate the degree of distortion at the tissue-lead interface with measured electrical parameters. Instantaneous impedance values obtained during fixation serve as a superior indicator of an acceptable lead implantation, and should therefore be carefully monitored during implantation.
引用
收藏
页码:883 / 891
页数:9
相关论文
共 15 条
[1]   MEASUREMENT OF THE OHMIC ELECTRODE-TISSUE CONTACT RESISTANCE AND THE VOLTAGE APPLIED TO MYOCARDIUM [J].
BENZING, G ;
ILTIS, R ;
HELMSWORTH, JA ;
SCHREIBER, T ;
KAPLAN, S .
JOURNAL OF BIOMEDICAL ENGINEERING, 1980, 2 (01) :3-8
[2]   Isolated four-chamber working swine heart model [J].
Chinchoy, E ;
Soule, CL ;
Houlton, AJ ;
Gallagher, WJ ;
Hjelle, MA ;
Laske, TG ;
Morissette, J ;
Iaizzo, PA .
ANNALS OF THORACIC SURGERY, 2000, 70 (05) :1607-1614
[3]   Pacing impedance variability in tined steroid eluting leads [J].
Danilovic, D ;
Ohm, OJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (07) :1356-1363
[4]   Effects of high energy shocks on pacing impedance during transvenous ICD implantation [J].
HiddenLucet, F ;
Hartung, WM ;
McTeague, KK ;
Walter, PF ;
Langberg, JT .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (05) :752-757
[5]   In vitro studies of human hearts [J].
Hill, AJ ;
Laske, TG ;
Coles, JA ;
Sigg, DC ;
Skadsberg, ND ;
Vincent, SA ;
Soule, CL ;
Gallagher, WJ ;
Iaizzo, PA .
ANNALS OF THORACIC SURGERY, 2005, 79 (01) :168-177
[6]  
IRNICH W, 1975, ADV PACEMAKING TECHN, P247
[7]  
LASKE TG, 2002, IEEE T
[8]   TIME COURSE OF TRANSVENOUS PACEMAKER STIMULATION IMPEDANCE, CAPTURE THRESHOLD, AND ELECTROGRAM AMPLITUDE [J].
PLATIA, EV ;
BRINKER, JA .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1986, 9 (05) :620-625
[9]   INSULATION FAILURE WITH BIPOLAR POLYURETHANE PACING LEADS [J].
RAYMOND, RD ;
NANIAN, KB .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1984, 7 (03) :378-380
[10]   Low pacing impedance in pacemaker patients with cardiac diseases [J].
Schuchert, A ;
vanLangen, H ;
Michels, K ;
Meinertz, T .
DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 1996, 121 (34-35) :1046-1049