Determination of a calibration factor for the nondestructive assay of Guidant 32P brachytherapy sources

被引:11
作者
Collé, R [1 ]
Zimmerman, BE [1 ]
Soares, CG [1 ]
Coursey, BM [1 ]
机构
[1] NIST, Phys Lab, Gaithersburg, MD 20899 USA
关键词
D O I
10.1016/S0969-8043(98)00166-3
中图分类号
O61 [无机化学];
学科分类号
070301 [无机化学]; 081704 [应用化学];
摘要
A calibration factor ('dial setting') for the nondestructive assay of Guidant TiNi-encapsulated P-32 intravascular brachytherapy wire sources has been determined for measurements with the Capintec CRC-12 (sic. 'dose calibrator') ionization chamber. The calibration factor was derived from ionization current measurements with the CRC-12 followed by very quantitative, destructive assays of the 32P content in two sources. Published by Elsevier Science Ltd.