Background and Purpose: We performed an ex vivo study to measure the effects of various factors on the deflection angles of a new flexible ureteroscope with active primary and secondary deflections. Materials and Methods: Active primary and secondary deflections in different sequences were measured initially without working devices (baseline measures). Nine different working instruments were then placed in the working channel, and alterations in ureteroscope deflectability were measured. Results: Baseline measures showed excellent angles of deflections, consistent with the manufacturer's reporting. Among the wires tested, only the stiff wire had a negative (minor) impact on deflectability. There were similar minimal decreases of deflection with baskets. The 1.6F electrohydraulic lithotripsy (EHL) probe had minimal influence on deflectability compared with the 3.0F EHL probe. The 200-mum and 365-mum laser fibers significantly (>20degrees) decreased the angles of deflection. Conclusions: In spite of the changes in deflectability with almost all working instruments, there were no significant decreases in the angles with most of them. The sequence is important in obtaining optimal deflection.
引用
收藏
页码:733 / 736
页数:4
相关论文
共 4 条
[1]  
Bagley D H, 1987, Surg Endosc, V1, P119, DOI 10.1007/BF00312699
[2]   INTRARENAL ACCESS WITH THE FLEXIBLE URETEROPYELOSCOPE - EFFECTS OF ACTIVE AND PASSIVE TIP DEFLECTION [J].
BAGLEY, DH .
JOURNAL OF ENDOUROLOGY, 1993, 7 (03) :221-224
[3]   Flexible ureterorenoscopy for the treatment of lower pole calyx stones:: Influence of different lithotripsy probes and stone extraction tools on scope deflection and irrigation flow [J].
Michel, MS ;
Knoll, T ;
Ptaschnyk, T ;
Köhrmann, KU ;
Alken, P .
EUROPEAN UROLOGY, 2002, 41 (03) :312-316
[4]   Flexible endoscope deflectability: Changes using a variety of working instruments and laser fibers [J].
Poon, M ;
Beaghler, M ;
Baldwin, D .
JOURNAL OF ENDOUROLOGY, 1997, 11 (04) :247-249