Efficacy of multiple ring and coil electrode radiofrequency ablation catheters for the creation of long linear lesions in the atria

被引:6
作者
McRury, ID [1 ]
Haines, DE [1 ]
机构
[1] Univ Virginia, Hlth Sci Ctr, Dept Internal Med, Div Cardiovasc, Charlottesville, VA 22908 USA
关键词
D O I
10.1016/S1350-4533(98)00074-5
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
Atrial fibrillation is an arrhythmia that may potentially be treated by creating long linear lesions in the atria to create lines of electrical conduction block. While this has been performed with success with open-heart surgery, it has been proposed that a less invasive catheter-based approach could achieve similar success. Radiofrequency energy catheter ablation was performed in vivo with two novel electrode catheters. Each was an expanding loop design: one with 3 mm ring electrodes; and one with 12.5 mm coil electrodes. Power delivery was controlled automatically with temperature (70 degrees C target) feedback from thermistors embedded in each electrode. A total of 39 lines of ablation were created in the atria of Il normal dogs. The coil electrodes were more effective in creating lesions than the ring electrodes with a similar prevalence of transmurality (89% vs. 85%) but a higher prevalence of continuous transmurality (35% vs. 5%). Sequential electrode energy delivery was better than simultaneous multipolar delivery due to varying efficiencies of tissue heating. Inadequate heating was observed in 47% of simultaneous versus 1% of sequential multipolar deliveries, and excessive heating in 6% versus 1% of cases, respectively. It is feasible to create linear atrial lesions with an expanding loop electrode catheter. Catheters with coil electrodes are more effective than those with ring electrodes. In order to avoid coagulum formation and inefficient heating, sequential electrode energy delivery is preferable to multipolar delivery. (C) 1998 IPEM. Published by Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:551 / 557
页数:7
相关论文
共 26 条
[1]
AVITALL B, 1994, PACE, V17, P774
[2]
AVITALL B, 1994, J AM COLL CARDIOL, V484, P276
[3]
BABBS CF, 1982, MED INSTRUM, V16, P245
[4]
ASSESSMENT OF EFFECTS OF A RADIOFREQUENCY ENERGY-FIELD AND THERMISTOR LOCATION IN AN ELECTRODE CATHETER ON THE ACCURACY OF TEMPERATURE-MEASUREMENT [J].
BLOUIN, LT ;
MARCUS, FI ;
LAMPE, L .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (05) :807-813
[5]
NONRHEUMATIC ATRIAL-FIBRILLATION - RISK OF STROKE AND ROLE OF ANTITHROMBOTIC THERAPY [J].
CAIRNS, JA ;
CONNOLLY, SJ .
CIRCULATION, 1991, 84 (02) :469-481
[6]
TEMPERATURE MONITORING DURING RADIOFREQUENCY CATHETER ABLATION PROCEDURES USING CLOSED-LOOP CONTROL [J].
CALKINS, H ;
PRYSTOWSKY, E ;
CARLSON, M ;
KLEIN, LS ;
SAUL, JP ;
KAY, GN ;
DAILEY, S ;
EPSTEIN, A ;
PLUMB, V ;
WHARTON, M ;
KANTER, R ;
SORRENTINO, R ;
GREENFIELD, RA ;
WALSH, E ;
WALDO, A ;
BIBLO, L ;
JOHNSON, N ;
LIEBERMAN, R ;
ROSENBAUM, D ;
HUANG, SKS ;
MITTLEMAN, RS ;
WAGSHAL, AB ;
GREENE, T ;
LIEM, B ;
LAUER, M ;
SUNG, R ;
YOUNG, C ;
EVANS, J ;
CASE, C ;
HACKETT, K ;
MILES, W ;
ZIPES, D ;
LAWRENCE, J ;
TOMASELLI, G ;
BERGER, R .
CIRCULATION, 1994, 90 (03) :1279-1286
[7]
INCREASING CATHETER ABLATION LESION SIZE BY SIMULTANEOUS APPLICATION OF RADIOFREQUENCY CURRENT TO 2 ADJACENT SITES [J].
CHANG, RJ ;
STEVENSON, WG ;
SAXON, LA ;
PARKER, J .
AMERICAN HEART JOURNAL, 1993, 125 (05) :1276-1284
[8]
THE SURGICAL-TREATMENT OF ATRIAL-FIBRILLATION .3. DEVELOPMENT OF A DEFINITIVE SURGICAL-PROCEDURE [J].
COX, JL ;
SCHUESSLER, RB ;
DAGOSTINO, HJ ;
STONE, CM ;
CHANG, BC ;
CAIN, ME ;
CORR, PB ;
BOINEAU, JP .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1991, 101 (04) :569-583
[9]
Radiofrequency catheter ablation of the atria eliminates pacing-induced sustained atrial fibrillation and reduces connexin 43 in dogs [J].
Elvan, A ;
Huang, XD ;
Pressler, ML ;
Zipes, DP .
CIRCULATION, 1997, 96 (05) :1675-1685
[10]
GOTTLIEB GJ, 1981, AM J PATHOL, V103, P292