Calibration of Raman spectrometer instrument response function with luminescence standards: An update

被引:49
作者
Frost, KJ [1 ]
McCreery, RL [1 ]
机构
[1] Ohio State Univ, Dept Chem, Columbus, OH 43210 USA
关键词
Raman; intensity calibration; intensity standards; instrument response function;
D O I
10.1366/0003702981943121
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An error in a previously reported procedure for correcting Raman spectra for instrument response was corrected, and then the approach was tested against reported values of Raman cross sections. The method uses luminescent standards with known emission curves of intensity vs. Raman shift, expressed in polynomial form. After acquisition of an observed emission curve from the luminescent standard, the correction of observed Raman spectra is automatic. Agreement of corrected peak areas with those calculated from literature cross sections is 5-20%. This correspondence is approximately as good as the agreement between different reported values, and much better than that observed For uncorrected spectra. Due to variations in the luminescence of the materials, it is likely that these or similar Raman intensity standards will be calibrated by the manufacturer or an independent agency.
引用
收藏
页码:1614 / 1618
页数:5
相关论文
共 6 条
[1]  
ETZ E, COMMUNICATION
[2]  
*GAL SYST, 1996, US GUID GRAMS 32 VER, P244
[3]  
Nestor J. R., 1973, Journal of Raman Spectroscopy, V1, P305, DOI 10.1002/jrs.1250010309
[4]   Simplified calibration of instrument response function for Raman spectrometers based on luminescent intensity standards [J].
Ray, KG ;
McCreery, RL .
APPLIED SPECTROSCOPY, 1997, 51 (01) :108-116
[5]  
RUDIK JM, 1985, J CHEM PHYS, V82, P1732
[6]   MEASUREMENTS OF THE ABSOLUTE RAMAN CROSS-SECTIONS OF BENZENE [J].
SCHOMACKER, KT ;
DELANEY, JK ;
CHAMPION, PM .
JOURNAL OF CHEMICAL PHYSICS, 1986, 85 (08) :4240-4247