The use of reduced pressure to expand the capabilities of TGA-FTIR

被引:10
作者
Jackson, RS
Rager, A
机构
[1] Bruker Opt Inc, Billerica, MA 01821 USA
[2] Bruker Opt GMBH, D-76275 Ettlingen, Germany
关键词
TGA-FTIR; infrared spectroscopy; evolved gas analysis; reduced pressure;
D O I
10.1016/S0040-6031(00)00694-8
中图分类号
O414.1 [热力学];
学科分类号
摘要
The benefits of coupling a Fourier transform infrared (FTIR) spectrometer to a thermogravimetric analyzer (TGA) are well known. FTIR spectroscopy provides information about the evolved gases that can lead to a better understanding of the thermal behavior of the sample. The combination of the two techniques can give both qualitative and quantitative information about thermal decomposition and reaction mechanisms. One of the current limitations of TGA-FTIR is the temperature of the transfer line and the gas cell in the infrared spectrometer. The temperature of the interface to the FTIR cannot be as high as the temperature that can be achieved in the furnace, and this leads to possible trapping of high boiling compounds that are evolved from the sample. In this paper, a system is described for performing TGA-FTIR measurements at reduced pressures, which significantly improves the gas Row characteristics of the system for high boiling compounds. It is shown that reduced pressure also minimizes sample carryover, and reduces or eliminates trapping of material in the transfer line. Data are also presented that show that using the pressure as an experimental parameter can yield additional information about the structure of a sample, and about the thermal decomposition process. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:415 / 424
页数:10
相关论文
共 13 条
[1]  
[Anonymous], COMMUNICATION
[2]  
Budavari S, 1989, The merck index
[3]  
CODY CA, 1981, AM LAB JAN, P93
[4]   THE USE OF A SMALL, LOW-COST FT-IR SPECTROMETER AS A MOBILE ANALYZER FOR FLUID STREAMS [J].
COMPTON, DAC ;
MARKELOV, M ;
MITTLEMAN, ML ;
GRASSELLI, JG .
APPLIED SPECTROSCOPY, 1985, 39 (06) :909-915
[5]   SIMULTANEOUS TG-DTA MASS-SPECTROMETRY TO 1550-DEGREES [J].
EMMERICH, WD ;
KAISERSBERGER, E .
JOURNAL OF THERMAL ANALYSIS, 1979, 17 (01) :197-212
[6]   HYPHENATED ON-THE-SPOT TG/FT-IR TECHNIQUE [J].
JANSEN, JAJ ;
VANDERMAAS, JH ;
DEBOER, AP .
APPLIED SPECTROSCOPY, 1992, 46 (01) :88-92
[7]  
KAISERSBERGER E, 1985, THERMOCHIM ACTA, V85, P275
[8]   FOURIER-TRANSFORM INFRARED EVOLVED GAS-ANALYSIS (FT-IR-EGA) [J].
LEPHARDT, JO .
APPLIED SPECTROSCOPY REVIEWS, 1982, 18 (02) :265-303
[9]   Thermogravimetric investigation of volatility of dioctyl phthalate from plasticized poly(vinyl chloride) [J].
Mrklic, Z ;
Kovacic, T .
THERMOCHIMICA ACTA, 1998, 322 (02) :129-135
[10]   STUDY OF RECYCLABLE POLYMER AUTOMOBILE UNDERCOATINGS CONTAINING PVC USING TG/FTIR [J].
POST, E ;
RAHNER, S ;
MOHLER, H ;
RAGER, A .
THERMOCHIMICA ACTA, 1995, 263 :1-6