Fluid dynamic optimization of a ventricular assist device using particle image velocimetry

被引:22
作者
Mussivand, T [1 ]
Day, KD [1 ]
Naber, BC [1 ]
机构
[1] Univ Ottawa, Inst Heart, Cardiovasc Devices Div, Ottawa, ON K1Y 4E9, Canada
关键词
D O I
10.1097/00002480-199901000-00007
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Thrombus formation and resulting thromboembolism are major risks that can impede the widespread use of ventricular assist devices (VADs). Adverse flow patterns (turbulence and stasis) have been implicated in thrombogenesis. This study focuses on optimization of VAD geometry, port orientation, and fluid dynamics to reduce thrombus formation. Particle image velocimetry with cross-correlation was performed using Amberlite particles suspended in distilled water. The transparent VADs were illuminated by halogen lamps. Four different VADs were tested in an iterative approach toward optimization. A peak shear stress of 9,100 dynes/cm(2) was noted in the first configuration immediately after the end of systole at the outlet port. Modifications in chamber geometry, port diameters and orientation, and valve enclosure design yielded shear stresses in the two subsequent geometries of 5,100 dynes/cm(2) and 1,900 dynes/cm(2), respectively. For the third iteration, a region of stasis occurred during the transition between the inlet port and the blood chamber. Further modifications were implemented, including a reduction in port diameters and further smoothing of the port entry region. This eliminated stasis and yielded a maximum shear level of 4,100 dynes/cm(2). In conclusion, optimization was achieved through geometric modification of the VAD, thus minimizing adverse flow conditions.
引用
收藏
页码:25 / 31
页数:7
相关论文
共 26 条
[1]  
BACHER RP, 1970, J LAB CLIN MED, V76, P485
[2]  
Baldwin J T, 1990, ASAIO Trans, V36, pM274
[3]  
Berger S, 1974, Prog Hemost Thromb, V2, P273
[4]  
Bronzino J.D., 1995, The Biomedical Engineering Handbook
[5]  
CHANDRAN KB, 1991, ARTIF ORGANS, V15, P420
[6]   ON THE HEMOLYTIC AND THROMBOGENIC POTENTIAL OF OCCLUDER PROSTHETIC HEART-VALVES FROM INVITRO MEASUREMENTS [J].
FIGLIOLA, RS ;
MUELLER, TJ .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1981, 103 (02) :83-90
[7]  
FORSTROM RJ, 1970, NEW ENGL J MED, V283, P208
[8]  
Guyton A C., TXB MED PHYSL
[9]  
Hans D, 1993, Eur J Med, V2, P157
[10]   Evolution of an electrohydraulic ventricular assist device through in vivo testing [J].
Hendry, PJ ;
Masters, RG ;
Keaney, M ;
Bourke, M ;
Mussivand, T ;
Keon, WJ .
ASAIO JOURNAL, 1996, 42 (05) :M350-M354