Heat Production, Nerve Function, and Morphology following Nerve Close Dissection with Surgical Instruments

被引:38
作者
Carlander, Johan [1 ,2 ]
Koch, C. [3 ]
Brudin, L. [4 ]
Nordborg, C. [5 ]
Gimm, O. [6 ]
Johansson, K. [1 ,6 ]
机构
[1] Linkoping Univ, Dept Clin & Expt Med, Fac Hlth Sci, Div Surg, Linkoping, Sweden
[2] Cent Lasarettet, Vasteras, Sweden
[3] Phys Tech Bundesanstalt, Braunschweig, Germany
[4] Kalmar Hosp, Dept Clin Physiol, Kalmar, Sweden
[5] Sahlgrens Univ Hosp, Dept Pathol, Gothenburg, Sweden
[6] Linkoping Univ, Cty Council Ostergotland, Dept Surg, Linkoping, Sweden
关键词
LATERAL THERMAL-DAMAGE; TISSUE; MONOPOLAR; SCALPEL; ENERGY;
D O I
10.1007/s00268-012-1471-x
中图分类号
R61 [外科手术学];
学科分类号
100210 [外科学];
摘要
The aim of the present study was to compare an ultrasonically activated instrument (US), monopolar electrosurgery, and bipolar electrosurgery (ES) with respect to heat production, nerve function, and nerve morphology following in vivo application. The biceps femoris muscle of anesthetized rats was cut in a standardized manner longitudinally 1 mm adjacent to the sciatic nerve using US shears, a monopolar ES knife, or a bipolar ES scissors. Activation time and temperature were recorded continuously within 1-4 mm of the activation site ipsilateral and contralateral to the nerve with two thermoelectric microsensors. Temperature rise and time delay of reaching the temperature maximum, as an expression of heat spread within tissue, maximum temperature, and thermal dose (equivalent time of exposure at 43A degrees C) were measured and calculated. A total of 49 functional experiments were conducted. The electromyographic (EMG) potential was recorded distally. Nerve dysfunction was defined as more than 10% loss of the evoked EMG amplitude. Forty-eight nerves were coded and submitted to blind histopathological examination, and morphological damage was graded on a 4-grade scale. The maximum temperature elevation and the thermal dose were significantly higher for the bipolar ES compared with the US instrument ( = 0.024, = 0.049), and with much less variation of results for the US instrument. The monopolar ES maximum temperature and thermal dose were lower, but a very large variation occurred, probably as a result of more random electrical spread to the ground electrode and muscle motion artifacts. Functional loss was least common in the US group-without being significant-compared to bipolar and monopolar ES. Moderate and severe morphological damage was significantly less common in the US group than in the monopolar ES group ( = 0.041). We found no statistically significant correlation between the highest temperatures and the degree of morphological damage or functional loss The temperature elevation depends strongly on the distance to the activated instrument. The bipolar ES scissors generates a higher maximum temperature and thermal dose with a greater variation in than the US. Functional loss and severe morphological damage were uncommon in all groups.
引用
收藏
页码:1361 / 1367
页数:7
相关论文
共 20 条
[1]
Feasibility study for robotic radical prostatectomy cautery-free neurovascular bundle preservation [J].
Ahlering, TE ;
Eichel, L ;
Chou, D ;
Skarecky, DW .
UROLOGY, 2005, 65 (05) :994-997
[2]
Real-time thermography during energized vessel sealing and dissection [J].
Campbell, PA ;
Cresswell, AB ;
Frank, TG ;
Cuschieri, A .
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2003, 17 (10) :1640-1645
[3]
Comparison of experimental nerve injury caused by ultrasonically activated scalpel and electrosurgery [J].
Carlander, J ;
Johansson, K ;
Lindström, S ;
Velin, ÅK ;
Jiang, CH ;
Nordborg, C .
BRITISH JOURNAL OF SURGERY, 2005, 92 (06) :772-777
[4]
Comparison of monopolar electrocoagulation, bipolar electrocoagulation, ultracision, and ligasure [J].
Diamantis, Theodore ;
Kontos, Michael ;
Arvelakis, Antonios ;
Syroukis, Spiridon ;
Koronarchis, Dimitris ;
Papalois, Apostolos ;
Agapitos, Emmanuel ;
Bastounis, Elias ;
Lazaris, Andreas C. .
SURGERY TODAY, 2006, 36 (10) :908-913
[5]
Duck F.A., 1991, Physical Properties of Tissue: A comprehensive Reference Book
[6]
Measurement of thermal and ultrasonic properties of some biological tissues [J].
El-Brawany, M.A. ;
Nassiri, D.K. ;
Terhaar, G. ;
Shaw, A. ;
Rivens, I. ;
Lozhken, K. .
Journal of Medical Engineering and Technology, 2009, 33 (03) :249-256
[7]
Emam TA, 2003, ANN SURG, V237, P186
[8]
The Pedicle Effect and Direct Coupling Delayed Thermal Injuries to the Bile Duct After Laparoscopic Cholecystectomy [J].
Humes, David J. ;
Ahmed, Irfan ;
Lobo, Dileep N. .
ARCHIVES OF SURGERY, 2010, 145 (01) :96-98
[9]
Determination of temperature elevation in tissue during the application of the harmonic scalpel [J].
Koch, C ;
Friedrich, T ;
Metternich, F ;
Tannapfel, A ;
Reimann, HP ;
Eichfeld, U .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2003, 29 (02) :301-309
[10]
Kunde D, 2003, J Obstet Gynaecol, V23, P347