High-burst-strength, feedback-controlled bipolar vessel sealing

被引:317
作者
Kennedy, JS [1 ]
Stranahan, PL [1 ]
Taylor, KD [1 ]
Chandler, JG [1 ]
机构
[1] Valleylab Inc, Boulder, CO 80301 USA
来源
SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES | 1998年 / 12卷 / 06期
关键词
vessel sealing; bipolar coagulation; ultrasonic; hemostasis;
D O I
10.1007/s004649900733
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background. The inherent tedium of intracorporeal knot tying has stimulated greater interest in energy-based and mechanical alternatives for hemostasis. Methods: Three hundred thirty-one arteries and veins were sealed by application of precisely controlled electrothermal energy and physical pressure, allowing for brief cooling in compression, in experimental animals and fresh abattoir vessels. These seals were compared for bursting strength with occlusions by ultrasonic and bipolar coagulation, surgical clips, and ligatures. Results: Ultrasonic and bipolar occlusions were significantly less likely to have burst strengths greater than 400 mmHg as compared with seals, clips, and ligatures (p < 0.001). Seal competence could be visually assessed by its translucence. Conclusions: Precise energy control with physical compression, including a brief cooldown, produces a distinctive, translucent seal of partially denatured protein that can typically be transected after a single application. These seals have bursting strengths comparable to those of clips and ligatures and resist dislodgement because they are intrinsic to the vessel wall structure.
引用
收藏
页码:876 / 878
页数:3
相关论文
共 5 条
[1]  
Amaral J F, 1994, Endosc Surg Allied Technol, V2, P181
[2]   DEVELOPMENT AND INITIAL APPLICATION OF A REAL-TIME THERMAL CONTROL-SYSTEM FOR LASER-TISSUE WELDING [J].
KLIOZE, SD ;
POPPAS, DP ;
ROOKE, CT ;
CHOMA, TJ ;
SCHLOSSBERG, SM .
JOURNAL OF UROLOGY, 1994, 152 (02) :744-748
[3]  
NELSON MT, 1992, ARCH SURG-CHICAGO, V127, P718
[4]  
SIGEL B, 1965, SURG GYNECOL OBSTETR, V121, P823
[5]   CHANGES IN BIREFRINGENCE AS MARKERS OF THERMAL-DAMAGE IN TISSUES [J].
THOMSEN, S ;
PEARCE, JA ;
CHEONG, WF .