Performance of the platelet function analyser PFA-100® in testing abnormalities of primary haemostasis

被引:149
作者
Harrison, P
Robinson, MSC
Mackie, IJ
Joseph, J
McDonald, SJ
Liesner, R
Savidge, GF
Pasi, J
Machin, SJ
机构
[1] UCL, Dept Haematol, Haemostasis Res Unit, London WC1E 6HX, England
[2] Great Ormond St Hosp Sick Children, Dept Haematol, London WC1N 3JH, England
[3] St Thomas Hosp, Haemophilia Ctr, London, England
[4] Royal Free Hosp, Haemophilia Ctr, London NW3 2QG, England
关键词
platelets; bleeding time; PFA-100 (R); platelet disorders; von Willebrand's disease;
D O I
10.1097/00001721-199901000-00004
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The PFA-100(R) device is a new instrument for the in-vitro testing of platelet function. Primary haemostasis is stimulated by recording the closure time taken for platelets to seal a 150 mu m aperture in the centre of a membrane coated with collagen and either epinephrine or ADP, Patients with type 3 von Willebrand's disease (n = 4) all had infinitely prolonged closure times (> 200 s) with both types of cartridge. A patient with afibrinogenemia exhibited only slightly prolonged closure times of 111 and 166 s for the ADP and epinephrine membranes, respectively. Patients with Glanzmann's thrombasthenia (n = 6) and Bernard Soulier syndrome (n = 2) had grossly prolonged closure times (> 200 s) with both types of cartridges. These results confirmed that the PFA-100(R) system was highly dependent on normal von Willebrand factor, glycoprotein Ib and glycoprotein IIb/IIIa levels but not on plasma fibrinogen. Patients with storage pool disease (n = 6) and Hermansky Pudlak syndrome (n = 7) had prolonged closure times with the epinephrine cartridge. There was no evidence of enhanced platelet function in patients with antiphospholipid syndrome, in sickle-cell disease or thalassemia. However, ingestion of aspirin resulted in a near consistent and significant prolongation of the closure time for the epinephrine cartridge but not for the ADP cartridge in both normal subjects and patients. The test offers a reliable, reproducible, rapid and simple means of assessing high-shear platelet function in vitro. (C) 1999 Lippincott Williams & Wilkins.
引用
收藏
页码:25 / 31
页数:7
相关论文
共 11 条
  • [1] THE PRIMARY ANTIPHOSPHOLIPID SYNDROME - MAJOR CLINICAL AND SEROLOGICAL FEATURES
    ASHERSON, RA
    KHAMASHTA, MA
    ORDIROS, J
    DERKSEN, RHWM
    MACHIN, SJ
    BARQUINERO, J
    OUTT, HH
    HARRIS, EN
    VILARDELLTORRES, M
    HUGHES, GRV
    [J]. MEDICINE, 1989, 68 (06) : 366 - 374
  • [2] Carcao MD, 1998, BRIT J HAEMATOL, V101, P70
  • [3] Screening for von Willebrand disease with a new analyzer using high shear stress: A study of 60 cases
    Fressinaud, E
    Veyradier, A
    Truchaud, F
    Martin, I
    Boyer-Neumann, C
    Trossaert, M
    Meyer, D
    [J]. BLOOD, 1998, 91 (04) : 1325 - 1331
  • [4] Comparison of four commercial citrate blood collection systems for platelet function analysis by the PFA-100(TM) system
    Heilmann, EJ
    Kundu, SK
    Sio, R
    Garcia, C
    Gomez, R
    Christie, DJ
    [J]. THROMBOSIS RESEARCH, 1997, 87 (01) : 159 - 164
  • [5] KRATZER MAA, 1985, HAEMOSTASIS, V15, P357
  • [6] KUNDU SK, 1995, SEMIN THROMB HEMOST, V21, P106
  • [7] Characterization of an in vitro platelet function analyzer, PFA-100(TM)
    Kundu, SK
    Heilmann, EJ
    Sio, R
    Garcia, C
    Ostgaard, RA
    [J]. CLINICAL AND APPLIED THROMBOSIS-HEMOSTASIS, 1996, 2 (04) : 241 - 249
  • [8] MAMMEN EF, 1995, SEMIN THROMB HEMOST, V21, P113
  • [9] Mammen EF, 1998, SEMIN THROMB HEMOST, V24, P195, DOI 10.1055/s-2007-995840
  • [10] A comparison of the effects of aspirin on bleeding time measured using the Simplate(TM) method and closure time measured using the PFA-100(TM), in healthy volunteers
    Marshall, PW
    Williams, AJ
    Dixon, RM
    Growcott, JW
    Warburton, S
    Armstrong, J
    Moores, J
    [J]. BRITISH JOURNAL OF CLINICAL PHARMACOLOGY, 1997, 44 (02) : 151 - 155