Balloon catheter ablation to treat paroxysmal atrial fibrillation: What is the level of pulmonary venous isolation?

被引:93
作者
Reddy, Vivek Y. [1 ,2 ]
Neuzi, Petr [1 ]
d'AviLa, Andre [1 ,2 ]
Laragy, Margaret [1 ,2 ]
Malchano, Zachary J. [1 ,2 ]
Kralovec, Stepan [1 ]
Kim, Steven J. [3 ]
Ruskin, Jeremy N. [1 ,2 ]
机构
[1] Harvard Univ, Massachusetts Gen Hosp, Sch Med, Cardiac Arrhythmia Serv, Boston, MA 02114 USA
[2] Homolka Hosp, Dept Cardiol, Prague, Czech Republic
[3] St Jude Med Inc, Minnetonka, MN USA
关键词
catheter ablation; balloon catheter; cryoablation; laser; focused ultrasound; electroanatomical mapping; atria[ fibrillation;
D O I
10.1016/j.hrthm.2007.11.006
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND Unlike the initial balloon ablation catheters that were designed to deliver ablation Lesions within the pulmonary veins (PVs), the current balloon prototypes are fashioned to deliver lesions at the PV ostia. OBJECTIVE Using etectroanatomical mapping, this study evaluates the actual location of ablation Lesions generated by cryobased, laser-based, or uttrasound-based balloon catheters. METHODS In a total of 14 patients with paroxysmal atrial fibrillation, PV isolation was performed using either a cryoballoon catheter (8 patients), Laser catheter (4 patients) or ultrasound balloon catheter (2 patients). Patients underwent preprocedural computed tomographic/magnetic resonance imaging. An intracardiac ultrasound catheter was used to aid in positioning the balloon catheter at the PV ostium/antrum. In all patients, sinus rhythm bipolar voltage amplitude maps (using either CARTO with computed tomographic/magnetic resonance image integration or NavX mapping) were generated at baseline and after electrical PV isolation as confirmed by use of a circular mapping catheter. RESULTS Eectrical isolation was achieved in 100% of the PVs. Electroanatomical mapping revealed that after ablation with any of the 3 balloon catheters, the extent of isolation included the tubular portions of each PV to the Level of the PV ostia. However, the PV antral portions were Left Largely unablated with all 3 balloon technologies. CONCLUSION Using the current generation of balloon ablation catheters, electrical isolation occurs at the level of the PV ostia, but the antral regions are Largely unablated.
引用
收藏
页码:353 / 360
页数:8
相关论文
共 23 条
[1]   Three-dimensional analysis of pulmonary venous ostial and antral anatomy: Implications for balloon catheter-based pulmonary vein isolation [J].
Ahmed, J ;
Sohal, S ;
Malchano, ZJ ;
Holmvang, G ;
Ruskin, JN ;
Reddy, VY .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2006, 17 (03) :251-255
[2]   Ablation of atrial fibrillation in humans using a balloon-based ablation system: Identification of the site of phrenic nerve damage using pacing maneuvers and CARTO [J].
Antz, Matthias ;
Chun, Kr Julian ;
Ouyang, Feifan ;
Kuck, Karl-Heinz .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2006, 17 (11) :1242-1245
[3]   Initiation of atrial fibrillation by ectopic beats originating from the pulmonary veins - Electrophysiological characteristics, pharmacological responses, and effects of radiofrequency ablation [J].
Chen, SA ;
Hsieh, MH ;
Tai, CT ;
Tsai, CF ;
Prakash, VS ;
Yu, WC ;
Hsu, TL ;
Ding, YA ;
Chang, MS .
CIRCULATION, 1999, 100 (18) :1879-1886
[4]   Clinical outcome after radiofrequency catheter ablation of focal atrial fibrillation triggers [J].
Gerstenfeld, EP ;
Guerra, P ;
Sparks, PB ;
Hattori, K ;
Lesh, MD .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2001, 12 (08) :900-908
[5]   Catheter ablation of long-lasting persistent atrial fibrillation:: Critical structures for termination [J].
Haïssaguerre, M ;
Sanders, P ;
Hocini, M ;
Takahashi, Y ;
Rotter, M ;
Sacher, F ;
Rostock, T ;
Hsu, LF ;
Bordachar, P ;
Reuter, S ;
Roudaut, R ;
Clémenty, J ;
Jaïs, P .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2005, 16 (11) :1125-1137
[6]   Electrophysiological end point for catheter ablation of atrial fibrillation initiated from multiple pulmonary venous foci [J].
Haïssaguerre, M ;
Jaïs, P ;
Shah, DC ;
Garrigue, S ;
Takahashi, A ;
Lavergne, T ;
Hocini, M ;
Peng, JT ;
Roudaut, R ;
Clementy, J .
CIRCULATION, 2000, 101 (12) :1409-1417
[7]   Ablation therapy for atrial fibrillation (AF):: Past, present and future [J].
Jaïs, P ;
Weerasooriya, R ;
Shah, DC ;
Hocini, M ;
Macle, L ;
Choi, KJ ;
Scavee, C ;
Haïssaguerre, M ;
Clémenty, J .
CARDIOVASCULAR RESEARCH, 2002, 54 (02) :337-346
[8]   How to do circular mapping catheter-guided pulmonary vein antrum isolation: The Cleveland Clinic approach [J].
Kanj, Mohamed H. ;
Wazni, Oussama M. ;
Natale, Andrea .
HEART RHYTHM, 2006, 3 (07) :866-869
[9]   Pulmonary vein anatomy in patients undergoing catheter ablation of atrial fibrillation - Lessons learned by use of magnetic resonance imaging [J].
Kato, R ;
Lickfett, L ;
Meininger, G ;
Dickfeld, T ;
Wu, R ;
Juang, G ;
Angkeow, P ;
LaCorte, J ;
Bluemke, D ;
Berger, R ;
Halperin, HR ;
Calkins, H .
CIRCULATION, 2003, 107 (15) :2004-2010
[10]  
Keane David, 2002, Card Electrophysiol Rev, V6, P341, DOI 10.1023/A:1021115804352