The large Debye-Scherrer camera installed at SPring-8 BL02B2 for charge density studies
被引:430
作者:
Nishibori, E
Takata, M
Kato, K
Sakata, M
[1
]
Kubota, Y
Aoyagi, S
Kuroiwa, Y
Yamakata, M
Ikeda, N
机构:
[1] Nagoya Univ, Dept Appl Phys, Nagoya, Aichi 4648603, Japan
[2] Osaka Womens Univ, Osaka 5900035, Japan
[3] Okayama Univ, Dept Phys, Okayama 7008530, Japan
[4] JASRI, Kouto 6795198, Japan
来源:
关键词:
synchrotron radiation;
powder diffraction;
accurate structure analysis;
D O I:
10.1016/S0168-9002(01)00639-8
中图分类号:
TH7 [仪器、仪表];
学科分类号:
0804 ;
080401 ;
081102 ;
摘要:
The design and performance of a large Debye-Scherrer Camera with imaging plate (IP) as a detector. which was very recently installed at SPring-8, BL02B2, is reported. By taking advantage of high beam quality of SPring-8, the camera enables one a rapid collection of a high counting statistics and high angular resolution powder pattern, which can lead to accurate structure analyses. The camera also provides easy access to structural changes at varied temperatures between 15-1000 K. The camera provides a rapid and accurate powder diffraction system utilizing third generation SR. (C) 2001 Elsevier Science B.V. All rights reserved.