Fitting adsorption isotherms to the distribution data determined using packed micro-columns for high-performance liquid chromatography

被引:37
作者
Jandera, P
Bunceková, S
Mihlbachler, K
Guiochon, G
Backovská, V
Planeta, J
机构
[1] Univ Pardubice, Dept Analyt Chem, Pardubice 53210, Czech Republic
[2] Univ Tennessee, Dept Chem, Knoxville, TN 37996 USA
[3] Acad Sci Czech Republ, Inst Analyt Chem, CS-61142 Brno, Czech Republic
关键词
adsorption isotherms; benzophenone; o-cresol; phenol; mandelic acid; teicoplanin;
D O I
10.1016/S0021-9673(01)01002-0
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Knowing the adsorption isotherms of the components of a mixture on the chromatographic system used to separate them is necessary for a better understanding of the separation process and for the optimization of the production rate and costs in preparative high-performance liquid chromatography (HPLC). Currently, adsorption isotherms are usually measured by frontal analysis, using conventional analytical columns. Unfortunately, this approach requires relatively large quantities of pure compounds, and hence is expensive, especially in the case of pure enantiomers. In this work, we investigated the possible use of packed micro-bore and capillary HPLC columns for the determination of adsorption isotherms of benzophenone, o-cresol and phenol in reversed-phase systems and of the enantiomers of mandelic acid on a Teicoplanin chiral stationary phase. We found a reasonable agreement between the isotherm coefficients of the model compounds determined on micro-columns and on conventional analytical columns packed with the same material. Both frontal analysis and perturbation techniques could be used for this determination. The consumption of pure compounds needed to determine the isotherms decreases proportionally to the second power of the decrease in the column inner diameter, i.e. 10 times for a micro-bore column (1 mm I.D.) and 100 times for capillary columns (0.32 mm I.D.) with respect to 3.3 mm I.D. conventional columns. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:19 / 29
页数:11
相关论文
共 22 条
[1]   COVALENTLY BONDED TEICOPLANIN CHIRAL STATIONARY-PHASE FOR HPLC ENANTIOSEPARATIONS [J].
ARMSTRONG, DW ;
LIU, YB ;
EKBORGOTT, KH .
CHIRALITY, 1995, 7 (06) :474-497
[2]   Facile liquid chromatographic enantioresolution of native amino acids and peptides using a teicoplanin chiral stationary phase [J].
Berthod, A ;
Liu, YB ;
Bagwill, C ;
Armstrong, DW .
JOURNAL OF CHROMATOGRAPHY A, 1996, 731 (1-2) :123-137
[3]   Quantification of single solute and competitive adsorption isotherms using a closed-loop perturbation method [J].
Blümel, C ;
Hugo, P ;
Seidel-Morgenstern, A .
JOURNAL OF CHROMATOGRAPHY A, 1999, 865 (1-2) :51-71
[4]   POROUS CERAMIC BED SUPPORTS FOR FUSED-SILICA PACKED CAPILLARY COLUMNS USED IN LIQUID-CHROMATOGRAPHY [J].
CORTES, HJ ;
PFEIFFER, CD ;
RICHTER, BE ;
STEVENS, TS .
JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY & CHROMATOGRAPHY COMMUNICATIONS, 1987, 10 (08) :446-448
[5]   MESSUNG VON ADSORPTIONSISOTHERMEN AN KATALYSATOREN BEI HOHEN TEMPERATUREN MIT HILFE DER GAS-FESTKORPER-ELUIERUNGSCHROMATOGRAPHIE [J].
CREMER, E ;
HUBER, H .
ANGEWANDTE CHEMIE-INTERNATIONAL EDITION, 1961, 73 (13) :461-&
[6]   Adsorption isotherms of nonionic surfactants in SBA-15 measured by micro-column chromatography [J].
Findenegg, G. H. ;
Eltekov, A. Y. .
JOURNAL OF CHROMATOGRAPHY A, 2007, 1150 (1-2) :236-240
[7]  
HELFFERICH F, 1964, J CHEM EDUC, V41, P410
[8]  
Helfferich F G, 1970, MULTICOMPONENT CHROM
[9]   Design of the simulated moving bed process based on adsorption isotherm measurements using a perturbation method [J].
Heuer, C ;
Küsters, E ;
Plattner, T ;
Seidel-Morgenstern, A .
JOURNAL OF CHROMATOGRAPHY A, 1998, 827 (02) :175-191
[10]   DETERMINATION OF COMPETITIVE ADSORPTION-ISOTHERMS FOR MODELING LARGE-SCALE SEPARATIONS IN LIQUID-CHROMATOGRAPHY [J].
JACOBSON, JM ;
FRENZ, J .
JOURNAL OF CHROMATOGRAPHY, 1990, 499 :5-19