Numerical analysis of blood flow in the clearance regions of a continuous flow artificial heart pump

被引:32
作者
Anderson, J
Wood, HG
Allaire, PE
Olsen, DB
机构
[1] Univ Virginia, Charlottesville, VA 22903 USA
[2] Univ Utah, Artificial Heart Lab, Salt Lake City, UT USA
关键词
computational fluid dynamics; heart pump; hemolysis; thrombosis;
D O I
10.1046/j.1525-1594.2000.06580.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The CFVAD3 is the third prototype of a continuous flow ventricular assist device being developed for implantation in humans. The pump consists of a fully shrouded 4-blade impeller supported by magnetic bearings. On either side of this suspended rotating impeller is a small clearance region through which the blood flows. The spacing and geometry of these clearance regions are very important to the successful operation of this blood pump. Computational fluid dynamics (CFD) solutions for this flow were obtained using TascFlow, a software package available from AEA Technology, U.K. Flow in these clearance regions was studied parametrically by varying the size of the clearance, the blood flow rate into the pump, and the rotational speed of the pump. The numerical solutions yield the direction and magnitude of the flow and the dynamic pressure. Experimentally measured pump flow rates are compared to the numerical study. The results of the study provide guidance for improving pump efficiency. It is determined that current clearances can be significantly reduced to improve pump efficiency without negative impacts.
引用
收藏
页码:492 / 500
页数:9
相关论文
共 12 条
[1]   CENTRIFUGAL BLOOD PUMP WITH A MAGNETICALLY SUSPENDED IMPELLER [J].
AKAMATSU, T ;
NAKAZEKI, T ;
ITOH, H .
ARTIFICIAL ORGANS, 1992, 16 (03) :305-308
[2]   Design of a magnetic bearing-supported prototype centrifugal artificial heart Pump [J].
Allaire, PE ;
Maslen, EH ;
Kim, HC ;
Bearnson, GB ;
Olsen, DB .
TRIBOLOGY TRANSACTIONS, 1996, 39 (03) :663-669
[3]  
ANDERSON JB, UNPUB ARTIF ORGANS
[4]  
ANDERSON JB, IN PRESS ASAIO J
[5]  
ANDERSON JB, 1999, THESIS U VIRGINIA CH
[6]  
[Anonymous], 1985, Pump Handbook
[7]  
BELENGER J, 1983, ARTIF ORGANS, V1, P74
[8]   Medical progress - Implantable left ventricular assist devices [J].
Goldstein, DJ ;
Oz, MC ;
Rose, EA .
NEW ENGLAND JOURNAL OF MEDICINE, 1998, 339 (21) :1522-1533
[9]   Test controller design, implementation, and performance for a magnetic suspension continuous flow ventricular assist device [J].
Hilton, EF ;
Allaire, PE ;
Wei, NH ;
Baloh, MJ ;
Bearnson, G ;
Olsen, DB ;
Khanwilkar, P .
ARTIFICIAL ORGANS, 1999, 23 (08) :785-791
[10]   SYSTEM CONSIDERATIONS FAVORING ROTARY ARTIFICIAL HEARTS WITH BLOOD-IMMERSED BEARINGS [J].
JARVIK, RK .
ARTIFICIAL ORGANS, 1995, 19 (07) :565-570