HRRT Versus HR plus Human Brain PET Studies: An Interscanner Test-Retest Study

被引:59
作者
van Velden, Floris H. P. [1 ]
Kloet, Reina W. [1 ]
van Berckel, Bart N. M. [1 ]
Buijs, Fred L. [1 ]
Luurtsema, Gert [1 ]
Lammertsma, Adriaan A. [1 ]
Boellaard, Ronald [1 ]
机构
[1] Vrije Univ Amsterdam, Med Ctr, Dept Nucl Med & PET Res, NL-1007 MB Amsterdam, Netherlands
关键词
brain PET; high-resolution research tomograph; high spatial resolution; RESOLUTION RESEARCH TOMOGRAPH; PERFORMANCE EVALUATION; RECONSTRUCTION ALGORITHMS; DOPAMINE TRANSPORTER; GRAPHICAL ANALYSIS; RECEPTOR-BINDING; ECAT EXACT; QUANTIFICATION; SENSITIVITY; FLUMAZENIL;
D O I
10.2967/jnumed.108.058628
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
The high-resolution research tomograph (HRRT) is a dedicated human brain PET scanner. The purpose of this study was to compare the quantitative accuracy of the HRRT with that of the clinical HR+ PET scanner and to assess effects of differences in spatial resolution between both scanners (similar to 2.7 mm and similar to 7.0 mm for HRRT and HR+, respectively). Methods: Paired C-11-flumazenil scans of 7 healthy volunteers were assessed. For each volunteer, dynamic scans (including arterial sampling) were acquired on both scanners on the same day, thereby minimizing intersubject variability. Volume of distribution was generated using Logan plot analysis with plasma input. In addition, other plasma input, reference tissue (with pons as the reference tissue input), and parametric methods were included in the inter-scanner comparison. Results: Logan volume-of-distribution analysis of HRRT data showed higher values than that of HR+ data (slope with the intercept fixed at the origin of 1.14 +/- 0.10 to 1.19 +/- 0.10, depending on the HRRT reconstruction method used). Smoothing HRRT reconstructions with a 6-mm full width at half maximum gaussian kernel reduced this slope toward the line of identity (1.04 +/- 0.11 to 1.07 +/- 0.11), retaining good correlation between HR+ and HRRT data (r, similar to 0.98). Similar trends were observed for other plasma input, reference tissue, and parametric methods. However, after reference matching the reference tissue models showed lower HRRT kinetic parameter values than HR+ values (slope with fixed intercept, 0.90 +/- 0.10 to 0.94 +/- 0.13). Conclusion: Higher values of pharmacokinetic parameter values, obtained from HRRT versus HR+ PET studies, indicate improved HRRT PET quantification primarily due to a reduction in partial-volume effects.
引用
收藏
页码:693 / 702
页数:10
相关论文
共 35 条
[1]
Performance evaluation of the whole-body PET scanner ECAT EXACT HR+ following the IEC standard [J].
Adam, LE ;
Zaers, J ;
Ostertag, H ;
Trojan, H ;
Bellemann, ME ;
Brix, G .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1172-1179
[2]
[Anonymous], IEEE NUCL SCI S
[3]
Boellaard R, 2004, IEEE NUCL SCI CONF R, P2553
[4]
Evaluation of basis function and linear least squares methods for generating parametric blood flow images using 15O-water and positron emission tomography [J].
Boellaard, R ;
Knaapen, P ;
Rijbroek, A ;
Luurtsema, GJJ ;
Lammertsma, AA .
MOLECULAR IMAGING AND BIOLOGY, 2005, 7 (04) :273-285
[5]
Characteristics of a new fully programmable blood sampling device for monitoring blood radioactivity during PET [J].
Boellaard, R ;
van Lingen, A ;
van Balen, SCM ;
Hoving, BG ;
Lammertsma, AA .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 2001, 28 (01) :81-89
[6]
Boellaard R, 2001, J NUCL MED, V42, P808
[7]
PPET: A software tool for kinetic and parametric analyses of dynamic PET studies [J].
Boellaard, Ronald ;
Yaqub, M. ;
Lubberink, M. ;
Lammertsma, A. .
NEUROIMAGE, 2006, 31 :T62-T62
[8]
Brix G, 1997, J NUCL MED, V38, P1614
[9]
Byars LG, 2005, IEEE NUCL SCI CONF R, P2622
[10]
Performance evaluation of the ECAT HRRT: an LSO-LYSO double layer high resolution, high sensitivity scanner [J].
de Jong, Hugo W. A. M. ;
van Velden, Floris H. P. ;
Kloet, Reina W. ;
Buijs, Fred L. ;
Boellaard, Ronald ;
Lammertsma, Adriaan A. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (05) :1505-1526