Transmit SENSE

被引:614
作者
Katscher, U
Börnert, P
Leussler, C
van den Brink, JS
机构
[1] Philips Res Labs, Sector Tech Syst, D-22335 Hamburg, Germany
[2] Philips Med Syst, Best, Netherlands
关键词
spatially-selective RF pulses; 2D RF pulses; SENSE; parallel imaging;
D O I
10.1002/mrm.10353
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The idea of using parallel imaging to shorten the acquisition time by the simultaneous use of multiple receive coils can be adapted for the parallel transmission of a spatially-selective multidimensional RF pulse. As in data acquisition, a multidimensional RF pulse follows a certain k-space trajectory. Shortening this trajectory shortens the pulse duration. The use of multiple transmit coils, each with its own time-dependent waveform and spatial sensitivity, can compensate for the missing parts of the excitation k-space. This results in a maintained spatial definition of the pulse profile, while its duration is reduced. This work introduces the concept of parallel transmission with arbitrarily shaped transmit coils (termed "Transmit SENSE"). Results of numerical studies demonstrate the theoretical feasibility of the approach. The experimental proof of principle is provided on a commercial MR scanner. The lack of multiple independent transmit channels was addressed by combining the excitation patterns from two separate subexperiments with different transmit setups. Shortening multidimensional RF pulses could be an interesting means of making 3D RF pulses feasible even for fast T-2(*) relaxing species or strong main field inhomogeneities. Other applications might benefit from the ability of Transmit SENSE to improve the spatial resolution of the pulse profile while maintaining the transmit duration. (C) 2003 Wiley-Liss, Inc.
引用
收藏
页码:144 / 150
页数:7
相关论文
共 21 条
[1]  
Börnert P, 1998, MAGN RESON MATER PHY, V7, P166
[2]   Curved slice imaging [J].
Bornert, P ;
Schaffter, T .
MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (06) :932-939
[3]  
Bracewell R. N., 1986, FOURIER TRANSFORM IT, V31999
[4]   SPATIAL LOCALIZATION IN 2 DIMENSIONS USING NMR DESIGNER PULSES [J].
HARDY, CJ ;
CLINE, HE .
JOURNAL OF MAGNETIC RESONANCE, 1989, 82 (03) :647-654
[5]   BROAD-BAND NUCLEAR MAGNETIC-RESONANCE PULSES WITH TWO-DIMENSIONAL SPATIAL SELECTIVITY [J].
HARDY, CJ ;
CLINE, HE .
JOURNAL OF APPLIED PHYSICS, 1989, 66 (04) :1513-1516
[6]   Density compensation functions for spiral MRI [J].
Hoge, RD ;
Kwan, RKS ;
Pike, GB .
MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (01) :117-128
[7]  
KATSCHER U, 2002, P 10 ANN M ISMRM HAW, P189
[8]  
OHLIGER MA, 2001, P 9 ANN M ISMRM GLAS, P21
[9]   A LINEAR CLASS OF LARGE-TIP-ANGLE SELECTIVE EXCITATION PULSES [J].
PAULY, J ;
NISHIMURA, D ;
MACOVSKI, A .
JOURNAL OF MAGNETIC RESONANCE, 1989, 82 (03) :571-587
[10]   A K-SPACE ANALYSIS OF SMALL-TIP-ANGLE EXCITATION [J].
PAULY, J ;
NISHIMURA, D ;
MACOVSKI, A .
JOURNAL OF MAGNETIC RESONANCE, 1989, 81 (01) :43-56