We performed analytical validation of the new version of the Liaison N-Tact PTH (parathormone) assay according to NCCLS guidelines and compared this new generation of reagent with the Roche Elecsys PTH assay. We showed that this new version is a sensitive and precise method with good recovery. Significant carryover was observed, but with limited clinical implications. We demonstrated that the new version of the Liaison PTH is in reasonably good agreement with the Roche Elecsys PTH assay, and as we observed no differences in a hemodialyzed population, moving from one method to the other should not affect the daily follow-up of patients. However, one should be cautious with the highest values (> 500 pg/mL). We established reference intervals of 12-54 pg/mL for the Liaison and 14-52 pg/mL for the Elecsys assay.
引用
收藏
页码:105 / 107
页数:3
相关论文
共 7 条
[1]
Comparison of liaison N-tact PTH (diasorin) and N-tact PTH SPIRMA (diasorin) in hemodialyzed patients [J].
Cavalier, E ;
Delanaye, P ;
Krzesinski, JM ;
Chapelle, JP .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2005, 43 (08) :890-891
[2]
Evidence that the amino-terminal composition of non-(1-84) parathyroid hormone fragments starts before position 19 [J].
D'Amour, P ;
Brossard, JH ;
Räkel, A ;
Rousseau, L ;
Albert, C ;
Cantor, T .
CLINICAL CHEMISTRY, 2005, 51 (01) :169-176
[3]
On the evolving nature of understanding dialysis-related disorders [J].
Eknoyan, G ;
Lindberg, JS .
AMERICAN JOURNAL OF KIDNEY DISEASES, 2003, 41 (04) :S1-S3
[4]
GOCHMAN N, 1975, CLIN CHEM, V21, P1531
[5]
A more accurate method to estimate glomerular filtration rate from serum creatinine: A new prediction equation [J].
Levey, AS ;
Bosch, JP ;
Lewis, JB ;
Greene, T ;
Rogers, N ;
Roth, D .
ANNALS OF INTERNAL MEDICINE, 1999, 130 (06) :461-+
[6]
Inter-method variability in PTH measurement: Implication for the care of CKD patients [J].
Souberbielle, J-C ;
Boutten, A. ;
Carlier, M-C ;
Chevenne, D. ;
Coumaros, G. ;
Lawson-Body, E. ;
Massart, C. ;
Monge, M. ;
Myara, J. ;
Parent, X. ;
Plouvier, E. ;
Houillier, P. .
KIDNEY INTERNATIONAL, 2006, 70 (02) :345-350
[7]
Assay-specific decision limits for two new automated parathyroid hormone and 25-hydroxyvitamin D assays [J].
Souberbielle, JC ;
Fayol, V ;
Sault, C ;
Lawson-Body, E ;
Kahan, A ;
Cormier, C .
CLINICAL CHEMISTRY, 2005, 51 (02) :395-400