Quality assurance of X-ray spectrometry for chemical analysis

被引:28
作者
Barreiros, MA
Pinheiro, T
Araújo, MF
Costa, MM
Palha, M
da Silva, RC
机构
[1] INETI, P-1649038 Lisbon, Portugal
[2] ITN, P-2685953 Sacavem, Portugal
[3] Univ Nova Lisboa, FCT, P-2829516 Caparica, Portugal
[4] CFNUL, P-1649003 Lisbon, Portugal
关键词
X-ray spectrometry; quality assurance; quality control; traceability;
D O I
10.1016/S0584-8547(01)00285-3
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Three different techniques, energy dispersive X-ray fluorescence, total reflection X-ray fluorescence and particle induced X-ray emission were used to initiate an evaluation program on quality assurance (QA) procedures applied to X-ray spectrometry for chemical analysis. The use of standard methodologies to assure the statistical control of measurement data is the main objective of this work. Certified Reference Materials were used and up to 15 certified elements were analyzed to carry out the QA procedures. For the internal quality control, z-scores were calculated and control charts were produced. The plotted elemental data illustrate statistically controlled methodologies for the majority of the determinations. Even the cases where the control charts exhibit values out of control limits, the z-scores are below 3 in absolute value, indicating satisfactory results. Concerning external quality control the statistical methods applied showed that the results obtained for the three techniques are comparable, although some significant differences occur, mainly due to sample preparation. Therefore, the techniques are traceable to certified reference materials and the data gathered so far, enable to initiate a database for QA procedures. (C) 2001 Elsevier Science BN. All rights reserved.
引用
收藏
页码:2095 / 2106
页数:12
相关论文
共 18 条
[1]  
[Anonymous], 1995, GUID EXPR UNC MEAS
[2]  
[Anonymous], [No title captured]
[3]  
DZIUNIKOWSKI B, 1989, ENERGY DISPERSIVE XR, V24, P431
[4]   REFERENCE MATERIALS FOR QUALITY ASSURANCE IN SEA-WATER ANALYSIS - PERFORMANCE OF TOTAL-REFLECTION X-RAY-FLUORESCENCE IN THE INTERCOMPARISON AND CERTIFICATION STAGES [J].
FREIMANN, P ;
SCHMIDT, D ;
NEUBAUERZIEBARTH, A .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1993, 48 (02) :193-198
[5]  
*INT ORG STAND, 1996, 43 ISO INT ORG STAND
[6]  
*INT ORG STAND, 1991, 8258 ISO INT ORG STA
[7]  
*INT ORG STAND, 1994, 5725 ISO INT ORG STA
[8]  
Johansson S.A.E., 1995, PARTICLE INDUCED XRA, P451
[9]   TOTAL-REFLECTION X-RAY-FLUORESCENCE SPECTROSCOPY [J].
KLOCKENKAMPER, R ;
KNOTH, J ;
PRANGE, A ;
SCHWENKE, H .
ANALYTICAL CHEMISTRY, 1992, 64 (23) :A1115-+
[10]   APPLICABILITY OF PIXE TO THE ANALYSIS OF BIOLOGICAL REFERENCE MATERIALS [J].
MAENHAUT, W ;
VANDENHAUTE, J ;
DUFLOU, H .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1987, 326 (07) :736-738